FDA Adverse Event Malfunction Summary report: N

AXSYM RUBELLA IGG ANTIBODY

MDR report key: 1061573 · Received June 17, 2008

Report

Report Number
1415939-2008-00086
Event Type
Malfunction
Date Received
June 17, 2008
Date of Event
April 14, 2008
Report Date
June 2, 2008
Manufacturer
ABBOTT LABORATORIES
Product Code
LFX
PMA / PMN Number
K954045
Removal / Correction Number
1415939-6/06/08-004-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED ABBOTT TO REPORT THE INABILITY TO CALIBRATE A NEW LOT OF REAGENT FOR THE AXSYM RUBELLA IGG ASSAY AND OBSERVED ERROR CODE 1018 (CALIBRATOR A CHECK FAILURE). THE CUSTOMER STATED THEY WERE TROUBLESHOOTING AN OPTICS ISSUE AND WOULD WAIT UNTIL THE OPTICS ISSUE WAS RESOLVED BEFORE ATTEMPTING TO RESOLVE THE RUBELLA CALIBRATION ISSUE. ABBOTT FIELD SERVICE WAS SENT TO THE CUSTOMER FACILITY AND THE CUSTOMER REPORTED THE CALIBRATION ISSUE PERSISTED AFTER SERVICE. THE CUSTOMER WAS SENT A NEW LOT OF REAGENT AND UPON RECALIBRATION, STATED THE RUBELLA CALIBRATION ISSUE WAS UNSUCCESSFUL. ADDITIONALLY, THE CUSTOMER REPORTED THERE IS NO CALIBRATION ISSUE WITH OTHER AXSYM ASSAYS. ABBOTT REQUESTED THE CALIBRATION DATA FOR EVALUATION BY ABBOTT, AND SENT THE CUSTOMER A NEW LOT OF CALIBRATORS FOR TROUBLESHOOTING WHICH DID NOT RESOLVE THE ISSUE. THE CUSTOMER RETURNED THE REAGENTS AND CALIBRATORS FOR INVESTIGATION. THERE WAS NO IMPACT TO PT MANAGEMENT REPORTED BY THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AXSYM RUBELLA IGG ANTIBODY MEIA FOR THE MEASUREMENT OF IGM ANTIBODIES TO RUBELLA VIRUS LFX ABBOTT LABORATORIES NA 60943M100

Patients

Seq Age Sex Outcome Treatment
1 NI LOT# 58184M100| LOT# 61236M200| AXSYM RUBELLA CALIBRATORS:| AXSYM RUBELLA CALIBRATORS| AXSYM ANALYZER