TAPERED TOOL LEGEND, F2/8TA23
Report
- Report Number
- 1625507-2008-00039
- Event Type
- Malfunction
- Date Received
- June 17, 2008
- Report Date
- May 19, 2008
- Manufacturer
- MEDTRONIC POWERED SURGICAL SOLUTION
- Product Code
- HBB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- NURSE
Narratives
COMMENTS: MALFUNCTION WAS CONFIRMED ON EVAL. IT WAS NOTED THAT THE FRACTURE WAS CONSISTENT WITH EXCESSIVE BENDING FORCE APPLIED TO THE TOOL WHILE CUTTING. DISSECTING TOOL BREAKAGE IS A KNOWN PROBLEM THAT MAY OCCUR DURING USE OF HIGH SPEED DISSECTING TOOLS. EXCESSIVE CUTTING FORCE OR BENDING OF TOOLS DURING USE CAN CAUSE BREAKAGE. WARNING IS PROVIDED IN DEVICE USER MANUAL INDICATING "DO NOT USE EXCESSIVE PRESSURE, SUCH AS BENDING OR PRYING, ON ATTACHMENTS OR DISSECTING TOOLS. THIS MAY CAUSE TOOL TO BEND OR BREAK AND CAUSE INJURY TO PT, OPERATOR AND/OR OPERATING ROOM STAFF."
REPORT REC'D INDICATING THAT THE TOOL BROKE WHILE PERFORMING A CRANIOTOMY. THE BROKEN PIECE WAS IMMEDIATELY RETRIEVED, AND ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAPERED TOOL LEGEND, F2/8TA23 | PNEUMATIC SURG DRILL TOOL LEGEND | HBB | MEDTRONIC POWERED SURGICAL SOLUTION | NA | 5359S |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |