FDA Adverse Event Malfunction Summary report: N

TAPERED TOOL LEGEND, F2/8TA23

MDR report key: 1061520 · Received June 17, 2008

Report

Report Number
1625507-2008-00039
Event Type
Malfunction
Date Received
June 17, 2008
Report Date
May 19, 2008
Manufacturer
MEDTRONIC POWERED SURGICAL SOLUTION
Product Code
HBB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

COMMENTS: MALFUNCTION WAS CONFIRMED ON EVAL. IT WAS NOTED THAT THE FRACTURE WAS CONSISTENT WITH EXCESSIVE BENDING FORCE APPLIED TO THE TOOL WHILE CUTTING. DISSECTING TOOL BREAKAGE IS A KNOWN PROBLEM THAT MAY OCCUR DURING USE OF HIGH SPEED DISSECTING TOOLS. EXCESSIVE CUTTING FORCE OR BENDING OF TOOLS DURING USE CAN CAUSE BREAKAGE. WARNING IS PROVIDED IN DEVICE USER MANUAL INDICATING "DO NOT USE EXCESSIVE PRESSURE, SUCH AS BENDING OR PRYING, ON ATTACHMENTS OR DISSECTING TOOLS. THIS MAY CAUSE TOOL TO BEND OR BREAK AND CAUSE INJURY TO PT, OPERATOR AND/OR OPERATING ROOM STAFF."

Description of Event or Problem · 1

REPORT REC'D INDICATING THAT THE TOOL BROKE WHILE PERFORMING A CRANIOTOMY. THE BROKEN PIECE WAS IMMEDIATELY RETRIEVED, AND ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAPERED TOOL LEGEND, F2/8TA23 PNEUMATIC SURG DRILL TOOL LEGEND HBB MEDTRONIC POWERED SURGICAL SOLUTION NA 5359S

Patients

Seq Age Sex Outcome Treatment
1