FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM 3T

MDR report key: 10612595 · Received October 1, 2020

Report

Report Number
9611109-2020-00554
Event Type
Malfunction
Date Received
October 1, 2020
Date of Event
September 4, 2020
Report Date
September 30, 2020
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DWC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.10: THE AFFECTED DEVICE WAS RETURNED BACK TO THE MANUFACTURER SITE FOR INVESTIGATION. THE ROOT CAUSE CAUSE WAS A DAMAGED COMPRESSOR.

Description of Event or Problem · 0

SEE INITIAL REPORT.

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). THE SERVICE TECHNICIAN IN CHARGE VERIFIED THAT THE DEVICE TRIPPED THE MAINS POWER WHEN THE COOLING COMPRESSOR HAD STARTED POINTING TO A DEFECTIVE COMPRESSOR AS ROOT CAUSE OF THE REPORTED ISSUE. IN ADDITION A BUBBLING SOUND COULD BE HEARD FROM THE PATIENT TANK SUGGESTING A LEAK OF COOLANT. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

LIVANOVA (B)(4) RECEIVED A REPORT OF A HEATER-COOLER SYSTEM 3T SHUTDOWN DURING A SERVICE ACTIVITY. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1077339 HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND 16-02-80

Patients

Seq Age Sex Outcome Treatment
1