HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2020-00554
- Event Type
- Malfunction
- Date Received
- October 1, 2020
- Date of Event
- September 4, 2020
- Report Date
- September 30, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.10: THE AFFECTED DEVICE WAS RETURNED BACK TO THE MANUFACTURER SITE FOR INVESTIGATION. THE ROOT CAUSE CAUSE WAS A DAMAGED COMPRESSOR.
SEE INITIAL REPORT.
THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). THE SERVICE TECHNICIAN IN CHARGE VERIFIED THAT THE DEVICE TRIPPED THE MAINS POWER WHEN THE COOLING COMPRESSOR HAD STARTED POINTING TO A DEFECTIVE COMPRESSOR AS ROOT CAUSE OF THE REPORTED ISSUE. IN ADDITION A BUBBLING SOUND COULD BE HEARD FROM THE PATIENT TANK SUGGESTING A LEAK OF COOLANT. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
LIVANOVA (B)(4) RECEIVED A REPORT OF A HEATER-COOLER SYSTEM 3T SHUTDOWN DURING A SERVICE ACTIVITY. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1077339 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |