FDA Adverse Event
Malfunction
Summary report: N
MONACO
MDR report key: 1061127
·
Received June 13, 2008
Report
- Report Number
- 1937649-2008-00004
- Event Type
- Malfunction
- Date Received
- June 13, 2008
- Date of Event
- June 3, 2008
- Report Date
- June 16, 2008
- Manufacturer
- COMPUTERIZED MEDICAL SYSTEMS, AN ELEKTA COMPANY
- Product Code
- MUJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CUSTOMER ADVISORY FOR DETAILS OF THE ISSUE. THE ADVISORY WILL BE SENT TO ALL MONACO CUSTOMERS WORLDWIDE.
Description of Event or Problem · 1
WITH THE MONACO RADIATION TREATMENT PLANNING SYSTEM, IF THE USER SELECTS THE MONTE CARLO ALGORITHM FOR FINAL DOSE COMPUTATION, IT IS POSSIBLE THAT THE FINITE SIZE PENCIL BEAM (FSPB) ALGORITHM WILL BE USED INSTEAD. THERE IS NO INDICATION ON THE OUTPUT OF THE SYSTEM THAT THIS HAS OCCURRED WHICH MAY LEAD THE ONCOLOGIST TO MAKE A FALSE ASSUMPTION REGARDING THE ACCURACY OF THE DOSE PRESENTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONACO | RADIATION TREATMENT PLANNING SYSTEM | MUJ | COMPUTERIZED MEDICAL SYSTEMS, AN ELEKTA COMPANY | REL 1.0.0 - 1.0.2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |