FDA Adverse Event Malfunction Summary report: N

MONACO

MDR report key: 1061127 · Received June 13, 2008

Report

Report Number
1937649-2008-00004
Event Type
Malfunction
Date Received
June 13, 2008
Date of Event
June 3, 2008
Report Date
June 16, 2008
Manufacturer
COMPUTERIZED MEDICAL SYSTEMS, AN ELEKTA COMPANY
Product Code
MUJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER ADVISORY FOR DETAILS OF THE ISSUE. THE ADVISORY WILL BE SENT TO ALL MONACO CUSTOMERS WORLDWIDE.

Description of Event or Problem · 1

WITH THE MONACO RADIATION TREATMENT PLANNING SYSTEM, IF THE USER SELECTS THE MONTE CARLO ALGORITHM FOR FINAL DOSE COMPUTATION, IT IS POSSIBLE THAT THE FINITE SIZE PENCIL BEAM (FSPB) ALGORITHM WILL BE USED INSTEAD. THERE IS NO INDICATION ON THE OUTPUT OF THE SYSTEM THAT THIS HAS OCCURRED WHICH MAY LEAD THE ONCOLOGIST TO MAKE A FALSE ASSUMPTION REGARDING THE ACCURACY OF THE DOSE PRESENTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONACO RADIATION TREATMENT PLANNING SYSTEM MUJ COMPUTERIZED MEDICAL SYSTEMS, AN ELEKTA COMPANY REL 1.0.0 - 1.0.2