MICROSTAAR INJECTOR
Report
- Report Number
- 2023826-2008-00823
- Event Type
- Malfunction
- Date Received
- June 13, 2008
- Date of Event
- May 12, 2008
- Report Date
- May 22, 2008
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- KYB
- PMA / PMN Number
- K954600
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NURSE
Narratives
EVAL: CONCLUSION: - AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CAN NOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MFG OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.
IT WAS REPORTED THAT THE SURGEON INSERTED A AA4203TL SILICONE PLATE LENS AND IT WAS TORN DURING INSERTION. THE LENS WAS REMOVED WITHOUT ANY PT INJURY. THE REPORTER STATED THE INCIDENT WAS CAUSED BY THE INJECTOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSTAAR INJECTOR | INTRAOCULAR LENS GUIDE | KYB | STAAR SURGICAL CO. | MSI-TR | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | LENS MODEL AA4203TL| CARTRIDGE MODEL MTC- 60CFP |