SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE
Report
- Report Number
- 2023826-2008-00822
- Event Type
- Malfunction
- Date Received
- June 13, 2008
- Date of Event
- May 22, 2008
- Report Date
- May 22, 2008
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- P900048
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION RESULTS: OTHER: VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED THAT PIECES OF THE OPTIC AND A HAPTIC WAS TORN OFF AND MISSING AND THERE WAS A CLEAR SURGICAL RESIDUE ON THE LENS. CONCLUSION: OTHER - AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CANNOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.
IT WAS REPORTED THAT AFTER THE SURGEON INSERTED A THREE PIECE SILICONE LENS, AND THE LENS TORE UPON INSERTION. THE LENS WAS REMOVED WITHOUT ANY PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | AQ2010V | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | STAARVISC II| CARTRIDGE MODEL AQ CARTRIDGE-FP| INJECTOR MODEL MSI-TM UNKNOWN |