FDA Adverse Event Malfunction Summary report: N

MICROSTAAR FOAM TIP PLUNGER

MDR report key: 1061029 · Received June 13, 2008

Report

Report Number
2023826-2008-00817
Event Type
Malfunction
Date Received
June 13, 2008
Date of Event
May 28, 2008
Report Date
May 30, 2008
Manufacturer
STAAR SURGICAL
Product Code
MSS
PMA / PMN Number
K980696
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL: RESULTS - A VISUAL INSPECTION OF THE RETURNED PRODUCTS SHOWS THE FOAM TIP ON THE PLUNGER IS BROKEN, ONE HAPTIC IS TORN OFF OF THE LENS AND THE OPTIC IS TORN IN HALF. THE CARTRIDGE HAS NO VISIBLE DAMAGE. THE CARTRIDGE AND LENS HAS CLEAR SURGICAL RESIDUE ON THEM. IT SHOULD BE NOTED THAT THE INJECTOR WAS NOT RECEIVED FOR INSPECTION. CONCLUSIONS - AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAS BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

THE REPORTER STATED THAT AS THE SURGEON WAS INSERTING A +24.0 DIOPTER ONE PIECE COLLAMER LENS USING THE FTP INDIGO IN THE INDIGO-P INJECTOR AND THE PLUNGER ROD BENT AND TORE THE LENS. THE SURGEON CUT THE LENS TO REMOVE IT AND PUT IN ANOTHER SAME TYPE LENS, NO PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROSTAAR FOAM TIP PLUNGER INTRAOCULAR LENS FOLDERS AND INJECTORS MSS STAAR SURGICAL FTP-INDIGO 1233322

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE: MODEL SFC25FP LOT#1235958| LENS: MODEL CC4204BF| INJECTOR: MODEL INDIGO-P LOT#UNK