FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1061019 · Received June 13, 2008

Report

Report Number
2023826-2008-00808
Event Type
Malfunction
Date Received
June 13, 2008
Date of Event
May 16, 2008
Report Date
May 20, 2008
Manufacturer
STAAR SURGICAL
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL: RESULTS - A VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWS ONE HAPTIC IS TORN OFF INTO THE OPTIC AND IS MISSING. THERE IS CLEAR SURGICAL RESIDUE ON THE LENS. IT SHOULD BE NOTED THAT THE INJECTOR AND CARTRIDGE WERE NOT RETURNED FOR EVAL. CONCLUSIONS - A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH THE CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MANUFACTURING, RELEASE TESTING, AND EVAL OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

THE REPORTER STATED THAT THE SURGEON WAS INSERTING AN ASPHERIC THREE PIECE COLLAMER LENS MODEL CQ2015A AND THE HAPTIC TORE. THE SURGEON CUT THE LENS TO REMOVE IT WITH NO PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 INJECTOR: MODEL MSI-PM LOT#UNK| CARTRIDGE: MODEL CQCARTRIDGE-FP LOT#UNK