FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE FREEDOM
MDR report key: 1060998
·
Received June 16, 2008
Report
- Report Number
- 2954323-2008-02117
- Event Type
- Malfunction
- Date Received
- June 16, 2008
- Date of Event
- May 18, 2008
- Report Date
- June 16, 2008
- Manufacturer
- ABBOTT DIABETES CARE INC, USA
- Product Code
- LFR
- Removal / Correction Number
- 2954323-04/28/08-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
RETAIN SAMPLE TESTING ON THE COMPLAINANT STRIP LOT HAS CONFIRMED A MANUFACTURING ISSUE RESULTING IN ERROR 3 MESSAGES WHEN THESE STRIPS ARE USED WITH FREESTYLE FREEDOM METERS. TESTING CONCLUDED THAT NOT ALL STRIPS IN AFFECTED VIALS ARE IMPACTED. THIS ISSUE IS ASSOCIATED WITH FIELD CORRECTION.
Description of Event or Problem · 1
A CUSTOMER REPORTED RECEIVING AN ERROR 3 MESSAGE WHEN A TEST STRIP WAS INSERTED IN THEIR FREESTYLE FREEDOM METER. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM | BLOOD GLUCOSE MONITORING SYSTEM | LFR | ABBOTT DIABETES CARE INC, USA | NA | 0627030 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |