FDA Adverse Event Malfunction Summary report: N

MICROCHOICE ELITE HIGH SPEED DRILL

MDR report key: 1060997 · Received June 12, 2008

Report

Report Number
1017294-2008-00241
Event Type
Malfunction
Date Received
June 12, 2008
Date of Event
May 23, 2008
Report Date
May 23, 2008
Manufacturer
CONMED LINVATEC
Product Code
DZI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION RESULTS: CONMED LINVATEC RECEIVED THIS HANDPIECE FOR EVAL AND CONFIRMED OVERHEATING. DURING TESTING THIS HANDPIECE, IT FAILED TO PERFORM TO SPECIFICATION RELATED TO MULTIPLE WORN PARTS. A REVIEW OF REPAIR HISTORY FOR THIS UNIT FOUND IT OVERDUE FOR PREVENTIVE MAINTENANCE; LAST REPAIR (B) (4) 2006. THE CUSTOMER WILL BE CONTACTED REGARDING THESE FINDINGS. ASSOCIATED REPORT: 1017294-2008-00240. THE HANDPIECE INSTRUCTION MANUAL WARNS THE USER OF THE FOLLOWING: OVERHEATING MIGHT OCCUR IF THE HANDPIECE OR ACCESSORY BEARINGS ARE WORN OR ARE NOT KEPT CLEAN. CONTINUALLY CHECK ALL PARTS OF THE HANDPIECE AND ATTACHMENTS FOR OVERHEATING. DISCONTINUE USE AND RETURN THE EQUIPMENT FOR SERVICE AS NECESSARY. OVERHEATING CAN CAUSE SERIOUS INJURY TO THE PT OR OPERATING ROOM PERSONNEL. THE USER MANUAL ALSO INFORMS THE USER THAT THE SERVICE INTERVAL FOR THIS DEVICE IS EVERY 6 MONTHS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING USE OF THIS DRILL, IT GOT HOT. THE USER FELT THE HEAT IN THE HANDPIECE AND DISCONTINUED USE. ANOTHER DRILL WAS USED TO COMPETE THE SURGERY WITHOUT INJURY OR DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROCHOICE ELITE HIGH SPEED DRILL DRILL DZI CONMED LINVATEC NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK CATALOG#: 00502006000| MICROCHOICE BUR GUARD| LOT#: JUL07S