FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 1060994 · Received June 12, 2008

Report

Report Number
3004464228-2008-00104
Event Type
Malfunction
Date Received
June 12, 2008
Date of Event
May 13, 2008
Report Date
May 13, 2008
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE EVAL INDICATES THAT A RETAINER ALLOWED A COMPONENT TO MOVE RESULTING IN DAMAGE WHICH PREVENTED THE PLUNGER FROM ADVANCING. THE USER WOULD HAVE BEEN AWARE OF A PROBLEM DURING THE FILL PROCESS. THERE IS RESISTANCE FELT IN FILLING THE POD WITH INSULIN AND A DISTINCT "CRACKLING" NOISE RESULTING FROM STRIPPING THE THREADS OF THE COMPONENT WHEN OVER-PRESSURISED. THE USER IS INSTRUCTED IN THE USER GUIDE TO FREQUENTLY MONITOR THEIR BG LEVELS. BY FOLLOWING THESE RECOMMENDATIONS, THE USER WOULD BECOME AWARE OF HIGH BG LEVELS AND COULD USE ANOTHER DEVICE OR BACKUP THERAPY IF REQUIRED.

Description of Event or Problem · 1

CUSTOMER SAID HE REMOVED A POD AFTER 2.5 DAYS BECAUSE HIS BGS "WERE RUNNING BETWEEN 240 AND 340" FOR TWO DAYS, AND HE COULD NOT GET THEM TO COME DOWN, DESPITE REPEATED BOLUSES. HE SAID THAT WHEN HE FILLED THE POD, HE FELT "A LOT OF RESISTANCE." HE MENTIONED THAT HE WAS ABLE TO ACTIVATE A NEW POD SUCCESSFULLY AND HIS BG CAME DOWN. NO FURTHER INFO REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L11657

Patients

Seq Age Sex Outcome Treatment
1 Other