MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD
Report
- Report Number
- 1056600-2008-00201
- Event Type
- Malfunction
- Date Received
- June 12, 2008
- Date of Event
- May 22, 2008
- Report Date
- June 12, 2008
- Manufacturer
- MICRO TYPING SYSTEMS, INC.
- Product Code
- KSZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- UNKNOWN
Narratives
OCD WAS UNABLE TO CONFIRM THE CUSTOMER'S COMPLAINT. RETAINED TESTING AND BATCH RECORD REVIEW WERE WITHIN RELEASE SPECIFICATIONS. THE CUSTOMER INDICATED THAT THEY WERE USING SAMPLES FROM NEONATES TO ADULTS (B) (6) WITH VARYING IMMUNO-SUPPRESS DIAGNOSIS FOR TESTING. LABELING CAUTIONS THE USER THAT IN SOME PTS (E.G., NEWBORNS, ELDERLY OR IMMUNOCOMPROMISED PTS) THE EXPECTED ABO ANTIBODIES MAY BE WEAK OR MISSING. INCIDENT IS MOST LIKELY SAMPLE RELATED. DIFFERENCES OBSERVED BETWEEN TUBE AND GEL TEST METHODS ARE NOT UNCOMMON. INCIDENT IS ISOLATED. ADD'L COMPLAINTS WERE OPENED FOR LOT 1111307037-09, (B) (4), MDR# 1056600-2008-00199 AND LOT 121707037-03, (B) (4), MDR # 1056600-2008-00200. (B) (4).
THE CUSTOMER REPORTED THAT THEY OBSERVED DISCREPANT RESULTS BETWEEN THE FORWARD AND REVERSE GROUPING USING MTS A/B/D MONOCLONAL AN REVERSE GROUPING CARD LOT# 101007037-14. THE CUSTOMER REPEATED THE SAMPLES IN THE TUBE METHOD AND OBTAINED ACCEPTABLE RESULTS. THE CUSTOMER USED AN ALTERNATE LOTS OF GEL CARDS AND OBTAINED THE SAME DISCREPANT RESULTS WITH THE SAMPLES. ERRONEOUS RESULTS WERE NOT REPORTED. FALSE NEGATIVE TEST RESULTS CAN LEAD TO TRANSFUSION OF INCOMPATIBLE BLOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD | A/B/O & RH TYPE TESTING REAGENT | KSZ | MICRO TYPING SYSTEMS, INC. | NA | 101007037-14 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |