FDA Adverse Event Injury Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL

MDR report key: 1060623 · Received June 13, 2008

Report

Report Number
2023826-2008-00821
Event Type
Injury
Date Received
June 13, 2008
Date of Event
May 21, 2008
Report Date
May 21, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P900048
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
MEDICAL ASSISTANT

Narratives

Additional Manufacturer Narrative · 1

EVALUATION CODES: RESULTS (OTHER): VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED THAT PART OF THE OPTIC AND A HAPTIC WAS TORN OFF AND THERE WAS A CLEAR SURGICAL RESIDUE ON THE LENS. CONCLUSION (OTHER) - AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CAN NOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MFG OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED AN AQ2010V THREE PIECE SILICONE LENS AND A HAPTIC TORE UPON INSERTION. THE INCISION WAS ENLARGED TO REMOVE THE LENS AND A SUTURE CLOSED THE WOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL CO. AQ2010V NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention CARTRIDGE MODEL AQ CARTRIDGE-FP| INJECTOR MODEL MSI-TM