FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1060492 · Received June 13, 2008

Report

Report Number
3003464075-2008-00257
Event Type
Injury
Date Received
June 13, 2008
Date of Event
May 19, 2008
Report Date
May 19, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT COMPLETING RINSEBACK OF THE PT'S BLOOD DUE TO CLOTTING. THE OPERATOR FAILED TO CLAMP THE PT'S FISTULA NEEDLE FOR THE RINSEBACK PROCESS. EVALUATION OF THE RETURNED CARTRIDGE CONFIRMED A DIALYSATE LEAK AT THE BALANCE CHAMBER OF THE CARTRIDGE. THE EXACT CAUSE OF THE LEAK IS UNDER INVESTIGATION. THE USER'S GUIDE INCLUDES THE FOLLOWING WARNING, "THE NXSTAGE CYCLER MAY NOT SENSE SLOW FLUID OR BLOOD LEAKS RESULTING FROM LOOSE CONNECTIONS, FAULTY COMPONENTS, VENOUS ACCESS DISCONNECTION OR OTHER POTENTIAL CAUSES. LEAKING FLUIDS COULD LEAD TO BLOOD LOSS, INJURY, OR DEATH. LEAKING FLUIDS COULD ALSO CAUSE A PERSON TO SLIP OR FALL. ALWAYS TIGHTEN AND RECHECK PT VASCULAR ACCESS AND ALL FLUID LINES, CONNECTIONS AND CLAMPS. SECURE THE BLOOD ACCESS DEVICE TO THE PT. VISUALLY INSPECT THE SYSTEM PERIODICALLY FOR EVIDENCE OF LEAKS. TERMINATE TREATMENT IF LEAK CAN NOT BE RESOLVED." A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADD'L INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. A SYSTEM ALARM OCCURRED DURING A ROUTINE HEMODIALYSIS TREATMENT. A CLEAR FLUID LEAK WAS OBSERVED. RINSEBACK WAS NOT COMPLETED DUE TO CLOTTING OF THE CIRCUIT, RESULTING IN AN ESTIMATED BLOOD LOSS OF 150CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8027711

Patients

Seq Age Sex Outcome Treatment
1 63 YR Other