- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- LAY USER/PATIENT
Narratives
Additional Manufacturer Narrative · 1
EVAL SUMMARY: THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. OCCLUSION, DELIVERY AND ACCURACY TESTS WERE PERFORMED, THE DEVICE WAS FOUND TO PASS ALL OCCLUSION, DELIVERY AND ACCURACY TESTS. NO OPERATIONAL OR FUNCTIONAL FAILURE WAS DETECTED AND THE PRODUCT WAS WITHIN SPECIFICATION.
Description of Event or Problem · 1
INFO WAS RECEIVED THAT REPORTED A PT WAS HOSPITALIZED IN 2008, DUE TO AN INCIDENT OF HYPOGLYCEMIA. THE PT WAS BROUGHT TO THE HOSPITAL WHEN HIS BLOOD GLUCOSE WAS <10 MG/DL. HE WAS TREATED WITH GLUCOSE GEL. THE INSULIN PUMP WILL BE RETURNED FOR EVAL; TO ENSURE THAT IT IS FUNCTIONING CORRECTLY AND DID NOT CONTRIBUTE TO THE REPORTED INCIDENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COZMO INSULIN PUMP | INSULIN PUMP | LZG | SMITHS MEDICAL MD (FORMERLY DELTEC, INC) | 1700 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |