BD RAPID DETECTION OF SARS-COV-2 VERITOR
Report
- Report Number
- 1119779-2020-00426
- Event Type
- Malfunction
- Date Received
- September 29, 2020
- Date of Event
- September 2, 2020
- Report Date
- October 13, 2021
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- QKP
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D4: MEDICAL DEVICE LOT #: 0233975. D4: MEDICAL DEVICE EXPIRATION DATE: 2020-12-10. H4: DEVICE MANUFACTURE DATE: 2020-08-25. H6: INVESTIGATION SUMMARY: THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT ALLEGES FALSE POSITIVE RESULTS WHEN USING KIT RAPID DETECTION OF SARS-COV-2 VERITOR (MATERIAL # 256082), BATCH NUMBER 0233975. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE POSITIVE/DISCREPANT RESULTS COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. INITIATED. AN INVESTIGATION AND TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. THE REPORTED ISSUE WAS UNABLE TO BE CONFIRMED. HOWEVER, THERE IS A TREND AGAINST FALSE POSITIVE RESULTS. BD HAS INITIATED CAPA (CORRECTIVE AND PREVENTIVE ACTION) # 1878253 TO FURTHER INVESTIGATE. THE ROOT CAUSE COULD NOT BE IDENTIFIED. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS.
IT WAS REPORTED WHILE USING A BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY A FALSE POSITIVE RESULT WAS OBTAINED. A REPEAT TEST WAS PERFORMED USING A PCR TEST METHOD AND THE RESULT WAS NEGATIVE. THE PATIENT SELF-QUARANTINED AT HOME. THERE WAS NO REPORT OF PATIENT IMPACT. (B)(4).
(B)(4). MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).
IT WAS REPORTED WHILE USING A BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY A FALSE POSITIVE RESULT WAS OBTAINED. A REPEAT TEST WAS PERFORMED USING A PCR TEST METHOD AND THE RESULT WAS NEGATIVE. THE PATIENT SELF-QUARANTINED AT HOME. THERE WAS NO REPORT OF PATIENT IMPACT. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1065134 | BD RAPID DETECTION OF SARS-COV-2 VERITOR | CORONAVIRUS ANTIGEN DETECTION SYSTEM | QKP | BECTON, DICKINSON & CO. (SPARKS) | 0233975 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |