FDA Adverse Event Malfunction Summary report: N

AXSYM DIGOXIN II REAGENT

MDR report key: 1060023 · Received June 12, 2008

Report

Report Number
2623532-2008-00015
Event Type
Malfunction
Date Received
June 12, 2008
Date of Event
April 19, 2007
Report Date
May 22, 2008
Manufacturer
ABBOTT DIAGNOSTICS INTERNATIONAL, LTD.
Product Code
KXT
PMA / PMN Number
K953718
Removal / Correction Number
2623532-05/30/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

AN INVESTIGATION WAS CONDUCTED TO REVIEW CUSTOMERS COMPLAINTS REGARDING THE NEW AXSYM TDM/TRANSPLANT DISK (LIST #03D53-04) THAT WAS RELEASED IN OCTOBER OF 2006 WHICH DID NOT CONTAIN THE DIGOXIN II ASSAY FILE. THE ROOT CAUSE WAS A PROCEDURAL ERROR TO THE DOCUMENTATION. DIGOXIN II LABELING WAS INADEQUATELY ASSESS DURING THE ASSESSMENT OF IMPACT OF THE DESIGN CHANGE TO THE AXSYM TDM/TRANSPLANT DISK TO OTHER ASSAYS (IE DIGOXIN II). THE CUSTOMER COMMUNICATION LETTER INSTRUCTED THE CUSTOMER TO CONTINUE USING AXSYM TDM/TRANSPLANT DISK TO INSTALL THE DIGOXIN II ASSAY. THE AXSYM DIGOXIN II PACKAGE INSERT WAS REVISED IN 2008 TO ONLY LIST ASSAY DISK THAT CONTAIN THE AXSYM DIGOXIN II ASSAY FILE. A NEW AXSYM DIGOXIN II SPECIFIC ASSAY DISK IS ANTICIPATED TO BE AVAILABLE TO AXSYM DIGOXIN II CUSTOMERS BY 10/01/2008. THIS IS A FINAL REPORT.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THE AXSYM DIGOXIN II REAGENT PROTOCOL DOES NOT APPEAR WHEN AXSYM TDM DISKETTE, LIST # 03D53-04 IS INSTALLED. THE AXSYM DIGOXIN II REAGENT IS NOT MENTIONED IN THE AXSYM TDM DISKETTE INSTRUCTIONS AS WELL. THE CUSTOMER INSTALLED AN OLDER DISKETTE, LIST # 02C43-01 IN ORDER TO SOLVE THE ISSUE. THERE WAS NO IMPACT TO PT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AXSYM DIGOXIN II REAGENT MEIA FOR THE MEASUREMENT OF DIGOXIN KXT ABBOTT DIAGNOSTICS INTERNATIONAL, LTD. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 NI AXSYM ANALYZER