LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM
Report
- Report Number
- 2024601-2008-00314
- Event Type
- Malfunction
- Date Received
- June 11, 2008
- Date of Event
- May 8, 2008
- Report Date
- May 14, 2008
- Manufacturer
- COSTA RICA
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
TAPER TYPE II. THE PRODUCT ASSOCIATED WITH THIS REPORT HAS NOT YET BEEN RETURNED. BASED UPON THE IMPLANT DATE PROVIDED BY THE REPORTER, THE CONNECTOR TYPE IS ASSUMED TO BE A TAPER II. THE REPORTER OF THE EVENT WAS ASKED TO RETURN THE PRODUCT FOR ANALYSIS. VISUAL EXAMINATION MAY CONFIRM OR DETERMINE ANOTHER CONNECTOR TYPE ASSOCIATED WITH THIS REPORT. ALLERGAN HAS ALSO ASKED THE REPORTER TO SUPPLY THE SERIAL NUMBER. ALLERGAN HAS NOT RECEIVED THE PRODUCT AT THIS TIME. THEREFORE, NO ANALYSIS OR TESTING HAS BEEN DONE. DEVICE LABELING ADDRESSED THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT, OR THE CONNECTOR TUBING." "CAUTION: FAILURE TO CREATE A STABLE, SMOOTH PATH FOR THE ACCESS PORT TUBING, WITHOUT SHARP TURNS OR BENDS, CAN RESULT IN TUBING BREAKS AND LEAKAGE."
REPORTED AS A PORT LEAK. "IT WAS LEAKING FROM THE TUBING THAT ATTACHES TO THE PORT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM 10.0CM | DEVICE FOR TREATMENT OF MORBID OBESITY | LTI | COSTA RICA | NA | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |