FDA Adverse Event Malfunction Summary report: N

BD RAPID DETECTION OF SARS-COV-2 VERITOR

MDR report key: 10599942 · Received September 29, 2020

Report

Report Number
1119779-2020-00416
Event Type
Malfunction
Date Received
September 29, 2020
Date of Event
September 7, 2020
Report Date
October 14, 2021
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
QKP
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH ADDITIONAL INFORMATION: D.4. MEDICAL DEVICE LOT #: 0211199. D.4. MEDICAL DEVICE EXPIRATION DATE: 2020-12-10. H.4. DEVICE MANUFACTURE DATE: 2020-08-06.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT ALLEGES FALSE POSITIVE RESULTS WHEN USING KIT RAPID DETECTION OF SARS-COV-2 VERITOR (MATERIAL # 256082), BATCH NUMBER 0211199. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE POSITIVE/DISCREPANT RESULTS COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. INITIATED AN INVESTIGATION AND TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. THE REPORTED ISSUE WAS UNABLE TO BE CONFIRMED. HOWEVER, THERE IS A TREND AGAINST FALSE POSITIVE RESULTS. BD HAS INITIATED CAPA (CORRECTIVE AND PREVENTIVE ACTION) # 1878253 TO FURTHER INVESTIGATE. THE ROOT CAUSE COULD NOT BE IDENTIFIED. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY 3 FALSE POSITIVE RESULTS WERE OBTAINED. A REPEAT TEST WAS PERFORMED USING A PCR TEST METHOD AND NASOPHARYNGEAL SWAB, THE RESULT WAS NEGATIVE. THE CUSTOMER STATED THE STAFF TESTED WAS ASYMPTOMATIC. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THERE WAS NO REPORT OF PATIENT IMPACT. EUA # (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY 3 FALSE POSITIVE RESULTS WERE OBTAINED. A REPEAT TEST WAS PERFORMED USING A PCR TEST METHOD AND NASOPHARYNGEAL SWAB, THE RESULT WAS NEGATIVE. THE CUSTOMER STATED THE STAFF TESTED WAS ASYMPTOMATIC. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THERE WAS NO REPORT OF PATIENT IMPACT EUA # (B)(4).

Additional Manufacturer Narrative · 1

EUA #(B)(4). MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED WHILE USING BD RAPID DETECTION OF SARS-COV-2 VERITOR ASSAY 3 FALSE POSITIVE RESULTS WERE OBTAINED. A REPEAT TEST WAS PERFORMED USING A PCR TEST METHOD AND NASOPHARYNGEAL SWAB, THE RESULT WAS NEGATIVE. THE CUSTOMER STATED THE STAFF TESTED WAS ASYMPTOMATIC. THIS TEST IS NOT INTENDED FOR USE ON ASYMPTOMATIC PATIENTS AND WAS THEREFORE USED OFF LABEL. THERE WAS NO REPORT OF PATIENT IMPACT EUA #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1067515 BD RAPID DETECTION OF SARS-COV-2 VERITOR CORONAVIRUS ANTIGEN DETECTION SYSTEM QKP BECTON, DICKINSON & CO. (SPARKS) 0211199

Patients

Seq Age Sex Outcome Treatment
1 Other