UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2008-00184
- Event Type
- Other
- Date Received
- June 10, 2008
- Date of Event
- May 16, 2008
- Report Date
- June 10, 2008
- Manufacturer
- BECMAN COULTER, INC.
- Product Code
- MMI
- PMA / PMN Number
- K023764
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NO INFORMATION
Narratives
THE CUSTOMER COLLECTS SAMPLES IN BD PST TUBES AND CENTRIFUGES SAMPLES IN THE POWER PROCESSOR VIA AN AUTOMATION LINE. QC WAS CURRENTLY PERFORMING WITHIN SPECIFICATIONS AT THE TIME CUSTOMER TECHNICAL SERVICE (CTS) WAS CONTACTED. QC RECOVERED WITHIN RANGE PRIOR TO AND AFTER THE EVENT. A FIELD SERVICE ENGINEER (FSE) WAS ON-SITE IN 2008: THE FSE FOUND SALT BUILD-UP ON SAMPLE PUMP VALVE. THE FSE REPLACED SAMPLE PUMP. THE FSE PERFORMED DIAGNOSTIC TESTING WHICH PASSED. PERFORMED PRECISION TESTING USING THE LOW LEVEL QC WHICH MET SPECIFICATIONS. THE CUSTOMER RE-SPINS THE SAMPLES PRIOR TO REPEAT ANALYSIS AND THE CUSTOMER WAS WORKING WITH THE SITES BIOMEDICAL ENGINEER TO ADDRESS THEIR CENTRIFUGE PROCESS. A CLEAR ROOT CAUSE OF THIS EVENT HAS NOT BEEN DETERMINED TO DATE. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUSLY ELEVATED ACCUTNI RESULTS GENERATED BY THE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM FOR SEVERAL PATIENTS. CUSTOMER TESTED SEVERAL PATIENT SAMPLES FOR ACCUTNI AND OBTAINED ERRONEOUSLY ELEVATED RESULTS WHICH REPEATED AND RECOVERED IN THE NORMAL RANGE. CUSTOMER HAS NOT SUPPLIED EXACT PATIENT RESULTS TO DATE. ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER DID NOT REPORT PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | MMI | BECMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |