FDA Adverse Event Malfunction Summary report: N

NRFIT

MDR report key: 10589227 · Received September 25, 2020

Report

Report Number
3012307300-2020-09833
Event Type
Malfunction
Date Received
September 25, 2020
Date of Event
September 7, 2020
Report Date
December 28, 2020
Manufacturer
SMITHS MEDICAL ASD, INC.
Product Code
LHI
UDI-DI
10610586044113
PMA / PMN Number
K162219
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

TWO PICTURES OF A CADD EXTENSION SET WERE RECEIVED FOR EVALUATION. THE PICTURE SHOWED ONE CASSETTE PRODUCT AND A CASSETTE EXTENSION SET WITH A BROKEN FEMALE LUER. ONE CADD CASSETTE PRODUCT FROM PART NUMBER 21-7605-24 AND UNKNOWN LOT NUMBER AND ONE CASSETTE EXTENSION TUBE WAS RECEIVED. THE SAMPLE WAS VISUALLY INSPECTED, AT A DISTANCE OF 12? TO 24? AND NORMAL CONDITIONS OF ILLUMINATION. THE CASSETTE EXTENSION TUBE HAD A PIECE OF YELLOW CAP INSIDE OF THE FEMALE LUER . THE SAMPLE COULD NOT BE FUNCTIONALLY TESTED BECAUSE OF THIS CONDITION. A 21-7006-24 PRODUCT WAS USED TO PERFORM A YELLOW CAP ASSEMBLY MANUALLY; THE YELLOW CAP COULD NOT BREAK IN MANUAL ASSEMBLY EVEN WHEN APPLYING EXCESSIVE FORCE. THE ASSEMBLY BETWEEN THE FEMALE LUER AND YELLOW CAP WAS FIXED TO A PRESS TO MAKE IT POSSIBLE TO APPLY MORE FORCE IN ASSEMBLY. THE YELLOW CAP COULD BE BROKEN WHEN FIXED TO THE PRESS, HOWEVER THIS IS NOT NORMAL USE CONDITIONS FOR THE DISPOSITIVE. THE YELLOW CAP PRESENTED DEFORMATIONS AT THE MOMENT TO FIX TO THE PRESS. THE FAILURE MODE COULD NOT BE REPLICATED, THEREFORE THE ROOT CAUSE IS NOT ATTRIBUTABLE AT THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

THE LOT NUMBER IS CURRENTLY UNAVAILABLE.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED INDICATING THAT COMPLAINED OF PAIN AND IT'S POSSIBLE THAT MEDICATION HAS NOT BEEN PASSING THROUGH THE CADD EXTENSION SET. THERE WERE NO ADVERSE EVENTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1051591 NRFIT SET, ADMINISTRATION, INTRAVASCULAR LHI SMITHS MEDICAL ASD, INC. 21-7605-24 10610586044113

Patients

Seq Age Sex Outcome Treatment
1