FDA Adverse Event Malfunction Summary report: N

3 ML BD LUER-LOK LUER-LOK TIP

MDR report key: 10588471 · Received September 25, 2020

Report

Report Number
1213809-2020-00659
Event Type
Malfunction
Date Received
September 25, 2020
Date of Event
September 4, 2020
Report Date
October 7, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096573
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: TWO PHOTOS AND 126 3ML SYRINGES IN FULLY SEALED BLISTER PACKS WERE RECEIVED AND EVALUATED. 123 OF THE SYRINGES WERE COMPLETELY BLANK WITH NO VISIBLE PRINT AND THREE OF THE SYRINGES HAD A SINGLE INK RING NEAR THE MIDDLE OF THE BARREL WITH NO OTHER PRINT PRESENT. ALL THE SYRINGES WERE REJECTABLE PER PRODUCT SPECIFICATION. THE POTENTIAL ROOT CAUSE FOR THE MISSING PRINT DEFECT IS ASSOCIATED WITH THE EQUIPMENT FAILURE. THERE WAS LIKELY AN INK PUMP FAILURE THAT QUICKLY STOPPED THE INK FLOW, PREVENTING INK FROM BEING APPLIED TO THE BARRELS. DHR WAS PERFORMED. ALL VISUAL INSPECTIONS WERE PERFORMED AS PER REQUIREMENT WITH A REQUALIFICATION FOR MISSING PRINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT 200 3 ML BD LUER-LOK¿ LUER-LOK¿ TIP HAD NO SCALE MARKINGS. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 309657. BATCH NO. 0150246. IT WAS REPORTED THAT MULTIPLE SYRINGES WERE FOUND WITH NO SCALE MARKINGS.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. FDA DEVICE PROBLEM CODE: 1675, FDA PATIENT PROBLEM CODE: 2645.

Description of Event or Problem · 1

IT WAS REPORTED THAT 200 3 ML BD LUER-LOK¿ LUER-LOK¿ TIP HAD NO SCALE MARKINGS. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 309657 BATCH NO. 0150246. IT WAS REPORTED THAT MULTIPLE SYRINGES WERE FOUND WITH NO SCALE MARKINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1055363 3 ML BD LUER-LOK LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309657 0150246 30382903096573

Patients

Seq Age Sex Outcome Treatment
1 Other