SENTINEL CEREBRAL PROTECTION SYSTEM (US)
Report
- Report Number
- 2134265-2020-13210
- Event Type
- Malfunction
- Date Received
- September 25, 2020
- Date of Event
- August 1, 2020
- Report Date
- September 25, 2020
- Manufacturer
- CLARET MEDICAL, INC.
- Product Code
- PUM
- PMA / PMN Number
- DEN160043
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DATE OF EVENT: ESTIMATED. DEVICE EVALUATED BY MFR: THE UNIT WAS RETURNED IN A GENERIC PLASTIC BAG, OVERALL VISUAL REVISION DID NOT IDENTIFY FAILURES OR EVIDENCE THAT COULD BE LOST DUE TO DECONTAMINATION PROCESS. THE UNIT WAS RETURNED WITH A UNKNOWN GUIDEWIRE. THE PROXIMAL FILTER WAS RETURNED UN-SHEATHED. THE DISTAL FILTER WAS RETURNED UN-SHEATHED AND INVERTED WITH THE STRUT TUBE BROKEN. THE ARTICULATING SHEATH (ADS) WAS RELAXED. THE INNER MEMBER WAS RETURNED WITHOUT VISIBLE DAMAGES UNDER THE REAR HANDLE SHELLS. FLUSHING WAS ACHIEVED IN ALL FLUSH PORTS. MICROSCOPIC INSPECTION OF THE DEVICE REVEALED NO DAMAGES TO THE DISTAL FILTER MEMBER. THE LOCATION OF THE STRUT TUBE BREAK WAS CONFIRMED. A TEST GUIDE WIRE WAS FULLY INSERTED FOR FUNCTIONAL TESTING. THE PROXIMAL FILTER SHEATHED AND UN-SHEATHED USING THE PROXIMAL FILTER SLIDER (#1). THE ADS RESPONDED AS EXPECTED WHEN TURNING THE ARTICULATING KNOB (#2). THE DISTAL FILTER SLIDER (#3) WAS PULLED AND NO FRICTION WAS FELT; HOWEVER, SHEATHING OF THE DISTAL FILTER WAS STOPPED DUE TO THE STRUT TUBE DAMAGE.
REPORTABLE ON RETURNED DEVICE ANALYSIS COMPLETED 18 SEPT 2020. A SENTINEL CEREBRAL PROTECTION SYSTEM WAS RETURNED WITH NO REPORTED ALLEGATION. UPON ANALYSIS, A DISTAL FILTER BREAK WAS DISCOVERED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1054102 | SENTINEL CEREBRAL PROTECTION SYSTEM (US) | EMBOLIC PROTECTION DEVICE | PUM | CLARET MEDICAL, INC. | CMS15-10C-US |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |