BD SEDI-40
Report
- Report Number
- 2243072-2020-01501
- Event Type
- Malfunction
- Date Received
- September 25, 2020
- Date of Event
- September 7, 2020
- Report Date
- October 1, 2020
- Manufacturer
- BECTON DICKINSON
- Product Code
- JPH
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10 DEVICE AVAILABLE FOR EVAL YES, D.10 RETURNED TO MANUFACTURER ON: 2020-09-07 H.6. INVESTIGATION SUMMARY BD RECEIVED THE FAULTY INSTRUMENT FROM THE CUSTOMER FOR INVESTIGATION. THE INSTRUMENT WAS EVALUATED BY VISUAL EXAMINATION AND FUNCTIONAL TESTING AND THE SPROCKET THAT DRIVES THE TUBE HOLDER WAS FOUND TO BE LOOSE. THE SPROCKET WAS RE-SEATED AND THE INSTRUMENT WAS CONFIRMED TO MEET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED BEFORE USE THE BD SEDI-40 HAD HARDWARE / SOFTWARE MALFUNCTION FOR ESR INSTRUMENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ¿MIXER DOES NOT WORK ( IT IS "ON" IN A SETUP)¿
(B)(4). MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED BEFORE USE THE BD SEDI-40 HAD HARDWARE / SOFTWARE MALFUNCTION FOR ESR INSTRUMENT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ¿MIXER DOES NOT WORK ( IT IS "ON" IN A SETUP).¿
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1055240 | BD SEDI-40 | ESR INSTRUMENT | JPH | BECTON DICKINSON | 15-42008 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |