DISPOSABLE ENDOSCOPE VALVES SET
Report
- Report Number
- 3014749926-2020-00001
- Event Type
- Malfunction
- Date Received
- September 25, 2020
- Date of Event
- August 17, 2020
- Report Date
- September 18, 2020
- Manufacturer
- GA HEALTH COMPANY LIMITED
- Product Code
- OCX
- UDI-DI
- 04897106950225
- PMA / PMN Number
- K181509
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- 003
Narratives
NO PATIENT INJURY OR SAFETY ISSUE REPORTED. THE DISASSEMBLY OF SUCTION VALVE MAY CAUSE PROLONGED PROCEDURE. IT IS DETERMINED THAT SUCH IMPACT WILL NOT BE A MAJOR ISSUE IN PROCEDURE AND THEREFORE IMMEDIATE CORRECTIVE ACTION FOR ON FIELD PRODUCT IS NOT REQUIRED. ANALYSIS HAS ESTIMATED THE PROBABILITY OF DEVICE FAILURE TO BE LOW. ON 2020-11-03 GA HEALTH COMPANY LIMITED (HEREINAFTER REFERRED TO AS "GA HEALTH") BEGAN VOLUNTARILY RECALLING SUCTION VALVE WHICH IS ALSO INCLUDED IN DISPOSABLE ENDOSCOPE VALVES SET FROM OUR CUSTOMERS AND HOSPITALS.
THE SUCTION BUTTON IS REPORTED STICKY AND/OR BROKEN DURING OR AFTER THE PROCEDURE FROM THREE LOTS. THE EVENTS OCCURRED OUTSIDE THE US AND THE DEVICE IS NOT RETURNED FOR ANALYSIS. THERE IS NO PATIENT INJURY REPORTED.
THE SUCTION BUTTON IS REPORTED STICKY AND/OR BROKEN DURING OR AFTER THE PROCEDURE FROM THREE LOTS. THE EVENTS OCCURRED OUTSIDE THE US AND THE DEVICE IS NOT RETURNED FOR ANALYSIS. THERE IS NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1055129 | DISPOSABLE ENDOSCOPE VALVES SET | SUCTION VALVE | OCX | GA HEALTH COMPANY LIMITED | GAR004 | 20022682, 20022627, 20021025 | 04897106950225 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |