FDA Adverse Event Malfunction Summary report: N

THERMOCOOL SMART TOUCH ELECTROPHYSIOLOGY CATHETER

MDR report key: 10583729 · Received September 24, 2020

Report

Report Number
2029046-2020-01322
Event Type
Malfunction
Date Received
September 24, 2020
Date of Event
July 23, 2020
Report Date
August 28, 2020
Manufacturer
BIOSENSE WEBSTER INC
Product Code
LYB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INITIAL REPORTER PHONE: (B)(6). THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE EVALUATION DETAILS: THE DEVICE WAS VISUALLY INSPECTED, AND THE PEBAX WAS FOUND DAMAGED (HOLE) WITH FOREIGN MATERIAL INSIDE. THEN, MAGNETIC SENSOR FUNCTIONALITY WAS TESTED ON CARTO AND THE CATHETER WAS PROPERLY VISUALIZED AND NO ERRORS WERE OBSERVED. THEN, THE FORCE SENSOR FEATURE WAS TESTED, AND IT WAS WORKING PROPERLY, THE FORCE VALUES WERE OBSERVED WITHIN SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED AND NO INTERNAL ACTION RELATED TO THE REPORTED COMPLAINT WERE IDENTIFIED. THE CUSTOMER COMPLAINT REGARDING FORCE ISSUE COULD NOT BE DUPLICATED DURING THE PRODUCT INVESTIGATION. HOWEVER, IT WAS DETERMINED THAT THE FOREIGN MATERIAL INSIDE THE PEBAX AREA COULD BE RELATED TO THE REPORTED ISSUE. THE ROOT CAUSE OF THE PEBAX DAMAGE CANNOT BE RELATED TO THE MANUFACTURING PROCESS AS THERE IS EVIDENCE THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THE DAMAGE COULD BE RELATED TO THE HANDLING OF THE DEVICE DURING THE PROCEDURE, HOWEVER, THIS CANNOT BE CONCLUSIVELY DETERMINED. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

A PATIENT UNDERWENT AN ATRIAL FLUTTER (AFL) ABLATION PROCEDURE WITH A THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER FOR WHICH BIOSENSE WEBSTER¿S PRODUCT ANALYSIS LAB IDENTIFIED A HOLE IN THE PEBAX WITH FOREIGN MATERIAL INSIDE. DURING THE PROCEDURE, NO FORCE VALUES WERE BEING DISPLAYED. A SECOND CATHETER WAS USED TO COMPLETE THE OPERATION. THERE WAS NO REPORT OF AN ADVERSE EVENT FOR THE PATIENT. THE FORCE ISSUE DURING THE PROCEDURE IS NOT MDR REPORTABLE. HOWEVER, THE HOLE IN THE PEBAX WITH FOREIGN MATERIAL INSIDE IS MDR REPORTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1046139 THERMOCOOL SMART TOUCH ELECTROPHYSIOLOGY CATHETER SIMILAR DEVICE D133601, PMA # P030031/S053 LYB BIOSENSE WEBSTER INC 30307857M

Patients

Seq Age Sex Outcome Treatment
1