FDA Adverse Event Malfunction Summary report: N

TRUE METRIX AIR

MDR report key: 10579978 · Received September 24, 2020

Report

Report Number
1000113657-2020-00703
Event Type
Malfunction
Date Received
September 24, 2020
Date of Event
August 26, 2020
Report Date
October 7, 2020
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K150052
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 07-OCT-2020: H6: UPDATED FDA'S METHOD, RESULT, AND CONCLUSION CODES. H10: METER WAS RETURNED FOR EVALUATION; NO DEFECT WAS DETECTED. TEST STRIPS WERE RETURNED FOR EVALUATION; NO DEFECT WAS DETECTED.

Additional Manufacturer Narrative · 1

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 15-SEP-2020 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER AND STATED REPLACEMENT PRODUCT RESOLVED INITIAL CONCERN.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 191, 183, 199, 185 AND 201 MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT IS 120 MG/DL. THE CUSTOMER FELT WELL, AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULT OF 319 MG/DL USING TRUEMETRIX AIR METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE DINING ROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 12/18/2021 AND TEST STRIPS WERE OPENED IN (B)(6) 2020. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 191 MG/DL, DATE: (B)(6) 2020, TIME: 9:53 AM FASTING. RESULT 2: 183 MG/DL, DATE: (B)(6) 2020, TIME: 8:36 AM FASTING. RESULT 3: 199 MG/DL, DATE: (B)(6) 2020, TIME: 8:59 AM FASTING. RESULT 4: 185 MG/DL, DATE: (B)(6) 2020, TIME: 8:35 AM FASTING. RESULT 5: 201 MG/DL, DATE: (B)(6) 2020, TIME: 8:01 AM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1048526 TRUE METRIX AIR SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TRUE METRIX NFRS24/CS 50CT MG/DL MX4289S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1