FDA Adverse Event Injury Summary report: N

INION CPS BIOABSORBABLE MESH PLATES

MDR report key: 10579082 · Received September 24, 2020

Report

Report Number
3003407235-2020-00004
Event Type
Injury
Date Received
September 24, 2020
Date of Event
October 2, 2017
Report Date
September 23, 2020
Manufacturer
INION OY
Product Code
JEY
PMA / PMN Number
K010352
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INION CPS® MESH PLATE WAS USED IN ORBITAL FLOOR REPAIR FOR A (B)(6) YEAR OLD MAN. FLUID ACCUMULATION WAS NOTICED 2.5 MONTHS POST OPERATIVELY, AND 3.5CC OF FLUID WAS DRAINED OUT BY PERCUTANEOUS NEEDLE ASPIRATION. THE NEEDLE ASPIRATION WAS REPEATED 3 TIMES DURING 8 - 10 MONTHS POST OPERATIVELY. NO GROWTH OF BACTERIA HAS BEEN FOUND IN FLUID COLLECTION. THE NUMBER OF PATIENTS EXPERIENCING THIS TYPE OF TISSUE REACTION IS VERY LOW. OCCURRENCE RATE OF DEGRADATION RELATED TISSUE REACTIONS (INION CPS®) IS 0,15 % AS REPORTED TO INION. FLUID ACCUMULATION IS A KNOWN ADVERSE EVENT WITH THE BIOABSORBABLE IMPLANTS, ESPECIALLY IF METABOLISM/VASCULARITY IS LOCALLY REDUCED OR POOR. TISSUE REACTIONS SUCH AS THIS ARE RARE, BUT MAY OCCUR 1-2 YEARS POSTOPERATIVELY WHEN THE DEVICE STARTS TO DEGRADE, ESPECIALLY WHEN A PLATE IS IMPLANTED UNDER A THIN SOFT TISSUE LAYER. SUCH REACTION DOES NOT AFFECT BONE HEALING NEGATIVELY, OR CAUSE PERMANENT DAMAGE OR IMPAIRMENT. THE OCCURRENCE RATE OF DEGRADATION RELATED TISSUE REACTIONS HAS BEEN REPORTED TO BE VERY LOW WITH POLYLACTIC ACID BASED IMPLANTS, WHICH IS IN LINE WITH INION'S POST MARKET DATA.

Description of Event or Problem · 1

INION CPS® MESH PLATE WAS USED IN ORBITAL FLOOR REPAIR FOR A (B)(6) YEAR OLD MAN. FLUID ACCUMULATION WAS NOTICED 2.5 MONTHS POST OPERATIVELY, AND 3.5CC OF FLUID WAS DRAINED OUT BY PERCUTANEOUS NEEDLE ASPIRATION. THE NEEDLE ASPIRATION WAS REPEATED 3 TIMES DURING 8 TO 10 MONTHS POST OPERATIVELY. NO GROWTH OF BACTERIA HAS BEEN FOUND IN FLUID COLLECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1048106 INION CPS BIOABSORBABLE MESH PLATES BONA PLATE JEY INION OY UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 51 YR Required Intervention