FDA Adverse Event Malfunction Summary report: N

2.7 DEGREE STRAIGHT SAGITAL SAW

MDR report key: 10574524 · Received September 23, 2020

Report

Report Number
3005985723-2020-00303
Event Type
Malfunction
Date Received
September 23, 2020
Date of Event
August 31, 2020
Report Date
November 23, 2020
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486032111
PMA / PMN Number
K143752
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED ISSUE - 1) STRAIGHT SAW ATTACHMENT HAD OIL LEAKING FROM IT WHEN OPENING TRAYS PRE-OP. 2) BLUE SQUARE DRIVER: HANDLE MOVING INDEPENDENTLY FROM THE DRIVER. NO SURGICAL DELAY. CASE TYPE / APPLICATION: TKA PRODUCT INSPECTION ¿ VISUAL INSPECTION CONFIRMED THE EVENT. THERE IS OIL RESIDUE ON THE ATTACHMENT. SEE ATTACHED IMAGE. PRODUCT HISTORY REVIEW ¿ REVIEW OF THE DEVICE HISTORY RECORDS INDICATE 50 DEVICES WERE MANUFACTURED AND 50 WERE ACCEPTED (INCLUDING SERIAL (B)(6) INTO FINAL STOCK ON 01/16/2018 WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW ¿ A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 212186, LOT NUMBER 35011217 SHOWS 00 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSION ¿ THE EVENT WAS CONFIRMED VIA VISUAL INSPECTION. NO ADDITIONAL INVESTIGATION OR SPECIFIC ACTIONS ARE REQUIRED. IF ADDITIONAL INFORMATION IS RECEIVED, THEN THE COMPLAINT WILL BE REOPENED.

Description of Event or Problem · 0

1) STRAIGHT SAW ATTACHMENT HAD OIL LEAKING FROM IT WHEN OPENING TRAYS PRE-OP. 2) BLUE SQUARE DRIVER: HANDLE MOVING INDEPENDENTLY FROM THE DRIVER. NO SURGICAL DELAY. CASE TYPE / APPLICATION: TKA.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

STRAIGHT SAW ATTACHMENT HAD OIL LEAKING FROM IT WHEN OPENING TRAYS PRE-OP. BLUE SQUARE DRIVER: HANDLE MOVING INDEPENDENTLY FROM THE DRIVER. NO SURGICAL DELAY. CASE TYPE / APPLICATION: TKA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1036630 2.7 DEGREE STRAIGHT SAGITAL SAW ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO MAKO SURGICAL CORP. 212186 35011217 / 3502814 00848486032111

Patients

Seq Age Sex Outcome Treatment
1 63 YR Other