FDA Adverse Event Malfunction Summary report: N

SCALED DRILL 2.5MM, AO FITTING

MDR report key: 10573207 · Received September 23, 2020

Report

Report Number
0008031020-2020-02261
Event Type
Malfunction
Date Received
September 23, 2020
Date of Event
August 27, 2020
Report Date
November 24, 2020
Manufacturer
STRYKER GMBH
Product Code
LXH
UDI-DI
07613252719861
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT COULD NOT BE CONFIRMED, SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO OTHER EVIDENCE WAS PROVIDED. THE EVENT DESCRIPTION STATES THAT: "DRILL WAS BENDING AT EXTREME ANGLE PRIOR TO BREAKAGE." BASED ON THIS STATEMENT, THE LIKELY ROOT CAUSE CAN BE ATTRIBUTED TO BE USER RELATED. THE FAILURE WAS LIKELY CAUSED BY THE BENDING OF THE DRILL DURING USE UNDER A POWER TOOL, WHICH APPLIES SIGNIFICANT FORCES ON THE DEVICE. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION. IF THE DEVICE IS RETURNED OR IF ANY ADDITIONAL INFORMATION IS PROVIDED, THE INVESTIGATION WILL BE REASSESSED.

Description of Event or Problem · 0

AS REPORTED: "2.5 MM X 450MM DRILL BIT (703966) BROKE OFF DURING DRILLING IN THE PUBIC RAMUS AND WAS NOT RETRIEVED. LOT # (KU90298). DRILL WAS BENDING AT EXTREME ANGLE PRIOR TO BREAKAGE. THE DRILL BIT WAS NOT RETRIEVED BECAUSE IT WOULD HAVE DONE MORE HARM TO THE PATIENT ATTEMPTING TO RETRIEVE IT. IT WAS FULLY CONTAINED IN THE BONE. THERE WAS NO PATIENT IMPACT AND NO SURGICAL DELAY. THE PROCEDURE CONTINUED WITHOUT FURTHER INCIDENT AND CONCLUDED WITH A POSITIVE OUTCOME FOR BOTH SURGEON AND PATIENT."

Additional Manufacturer Narrative · 1

DEVICE WILL NOT BE RETURNED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

AS REPORTED: "2.5 MM X 450MM DRILL BIT (703966) BROKE OFF DURING DRILLING IN THE PUBIC RAMUS AND WAS NOT RETRIEVED. LOT # (KU90298). DRILL WAS BENDING AT EXTREME ANGLE PRIOR TO BREAKAGE. THE DRILL BIT WAS NOT RETRIEVED BECAUSE IT WOULD HAVE DONE MORE HARM TO THE PATIENT ATTEMPTING TO RETRIEVE IT. IT WAS FULLY CONTAINED IN THE BONE. THERE WAS NO PATIENT IMPACT AND NO SURGICAL DELAY. THE PROCEDURE CONTINUED WITHOUT FURTHER INCIDENT AND CONCLUDED WITH A POSITIVE OUTCOME FOR BOTH SURGEON AND PATIENT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1039680 SCALED DRILL 2.5MM, AO FITTING ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH STRYKER GMBH 703966 KU90298 07613252719861

Patients

Seq Age Sex Outcome Treatment
1