FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1057276 · Received June 5, 2008

Report

Report Number
2023826-2008-00792
Event Type
Malfunction
Date Received
June 5, 2008
Date of Event
May 12, 2008
Report Date
May 12, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

EVAL: JOS A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH THE CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MANUFACTURING, RELEASE TESTING, AND EVAL OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. THIS COMPLAINT FILE WILL BE CLOSED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED A CQ2015A THREE PIECE COLLAMER LENS AND THE LENS TORE UPON INSERTION. THE LENS WAS REMOVED WITHOUT ANY PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. CQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 UNK CARTRIDGE MODEL CQ CARTRIDGE-FP| INJECTOR MODEL MSI-TM LOT NUMBER UNK