FDA Adverse Event
Malfunction
Summary report: N
COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA
MDR report key: 1057276
·
Received June 5, 2008
Report
- Report Number
- 2023826-2008-00792
- Event Type
- Malfunction
- Date Received
- June 5, 2008
- Date of Event
- May 12, 2008
- Report Date
- May 12, 2008
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- P990013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
EVAL: JOS A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH THE CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MANUFACTURING, RELEASE TESTING, AND EVAL OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. THIS COMPLAINT FILE WILL BE CLOSED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE SURGEON INSERTED A CQ2015A THREE PIECE COLLAMER LENS AND THE LENS TORE UPON INSERTION. THE LENS WAS REMOVED WITHOUT ANY PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | CQ2015A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | CARTRIDGE MODEL CQ CARTRIDGE-FP| INJECTOR MODEL MSI-TM LOT NUMBER UNK |