FDA Adverse Event Malfunction Summary report: N

HIGH DEFINITION LCD MONITOR

MDR report key: 10570081 · Received September 22, 2020

Report

Report Number
8010047-2020-06741
Event Type
Malfunction
Date Received
September 22, 2020
Report Date
December 30, 2020
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FET
PMA / PMN Number
K102379
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION HAS BEEN COMPLETED. THE ROOT CAUSE COULD NOT BE CONCLUSIVELY DETERMINED. PROBABLE CAUSES THAT COULD HAVE LED TO THE REPORTED EVENT INCLUDE THAT THE BUTTON FOR SWITCHING THE INPUT SIGNAL WAS NOT SET CORRECTLY ACCORDING TO THE PROPOSED INFORMATION OR THAT THE IMAGE PROCESSING BOARD WAS TEMPORARY DEFECTIVE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT. ADDITIONAL INFORMATION RECEIVED IDENTIFIED THIS EVENT OCCURRED DURING THE BEGINNING OF AN UNSPECIFIED DIAGNOSTIC PROCEDURE. ALTHOUGH, OTHER DEVICES WERE INVOLVED IN THE EVENT, THE CUSTOMER WAS UNABLE TO PROVIDE THEM. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE AND THERE WAS A 10 MINUTE DELAY. THE MONITOR WAS INSPECTED AND NO ANOMALIES WERE OBSERVED. THE MONITOR CABLE WAS NOTED TO BE UNDAMAGED, CONNECTED PROPERLY AND SECURE. PER THE CUSTOMER, THE INSTRUCTIONS FOR USE ARE BEING FOLLOWED.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND THE REPORTED EVENT COULD NOT BE DUPLICATED. THE CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED AT THIS TIME AS ALL SIGNALS FUNCTIONED NORMALLY. IF ADDITIONAL INFORMATION BECOMES AVAILABLE AT A LATER TIME, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

IT WAS REPORTED NO VIDEO WAS BEING PRODUCED FROM THE UNIT. ADDITIONAL INFORMATION IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1033357 HIGH DEFINITION LCD MONITOR HIGH DEFINITION LCD MONITOR FET OLYMPUS MEDICAL SYSTEMS CORP. OEV262H

Patients

Seq Age Sex Outcome Treatment
1