BD 60ML SYRINGE LUER-LOK TIP
Report
- Report Number
- 1911916-2020-00886
- Event Type
- Malfunction
- Date Received
- September 22, 2020
- Date of Event
- May 11, 2020
- Report Date
- October 5, 2020
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- UDI-DI
- 30382903096535
- PMA / PMN Number
- K980987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHARMACIST
Narratives
THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTIONS: B.3. DATE OF EVENT: 2020-09-10. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. B.5. DESCRIBE EVENT OR PROBLEM: THE EVENT DESCRIPTION HAS BEEN UPDATED WITH A NEW NUMBER OF OCCURRENCES, 10. G.4. DATE RECEIVED BY MANUFACTURER: 2020-09-10.
H.6. INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 309653 AND LOT NUMBER 0031559. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT BD 60ML SYRINGE LUER-LOK¿ TIP LEAKED PAST THE STOPPER. THIS OCCURRED ON 10 OCCASIONS DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 309653 BATCH NO. 0031559. IT WAS REPORTED THAT MEDICATION LEAKED PAST RUBBER STOPPER. EVENT DESCRIPTION PER EMAIL STATES: BD 50ML SYRINGE, LOT NUMBER 0031559, EXP DATE 01/31/2025, HAS HAD SEVERAL SYRINGES LEAKING FROM THE RUBBER STOPPER INTO THE BARREL OF THE SYRINGE THESE SYRINGES WERE USED TO PREPARE CHEMOTHERAPY, WHICH COULD HAVE EXPOSED PHARMACY TECHNICIAN TO CHEMOTHERAPY DURING THE PREPARATION PROCESS. TECHNICIAN HAS NOTED 4 TO 5 SYRINGES WITH THE SAME PROBLEM, ALL FROM THE SAME LOT NUMBER FDA SAFETY REPORT ID# (B)(4).
IT WAS REPORTED THAT BD 60ML SYRINGE LUER-LOK¿ TIP LEAKED PAST THE STOPPER. THIS OCCURRED ON 10 OCCASIONS DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 309653 BATCH NO. 0031559 IT WAS REPORTED THAT MEDICATION LEAKED PAST RUBBER STOPPER. EVENT DESCRIPTION PER EMAIL STATES: BD 50ML SYRINGE, LOT NUMBER 0031559, EXP DATE 01/31/2025, HAS HAD SEVERAL SYRINGES LEAKING FROM THE RUBBER STOPPER INTO THE BARREL OF THE SYRINGE THESE SYRINGES WERE USED TO PREPARE CHEMOTHERAPY, WHICH COULD HAVE EXPOSED PHARMACY TECHNICIAN TO CHEMOTHERAPY DURING THE PREPARATION PROCESS. TECHNICIAN HAS NOTED 4 TO 5 SYRINGES WITH THE SAME PROBLEM, ALL FROM THE SAME LOT NUMBER FDA SAFETY REPORT ID# (B)(4).
THE INITIAL REPORTER ALSO NOTIFIED THE FDA ON 19 MAY, 2020. MEDWATCH REPORT # MW5094483 REPORT SOURCE OTHER: MEDWATCH REPORT. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT BD 60ML SYRINGE LUER-LOK¿ TIP LEAKED PAST THE STOPPER. THIS OCCURRED ON 5 OCCASIONS DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 309653 BATCH NO. 0031559. IT WAS REPORTED THAT MEDICATION LEAKED PAST RUBBER STOPPER. EVENT DESCRIPTION PER EMAIL STATES: BD 50ML SYRINGE, LOT NUMBER 0031559, EXP DATE 01/31/2025, HAS HAD SEVERAL SYRINGES LEAKING FROM THE RUBBER STOPPER INTO THE BARREL OF THE SYRINGE THESE SYRINGES WERE USED TO PREPARE CHEMOTHERAPY, WHICH COULD HAVE EXPOSED PHARMACY TECHNICIAN TO CHEMOTHERAPY DURING THE PREPARATION PROCESS. TECHNICIAN HAS NOTED 4 TO 5 SYRINGES WITH THE SAME PROBLEM, ALL FROM THE SAME LOT NUMBER FDA SAFETY REPORT ID# (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1035684 | BD 60ML SYRINGE LUER-LOK TIP | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 309653 | 0031559 | 30382903096535 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |