FDA Adverse Event Malfunction Summary report: N

FULL FACE MASK

MDR report key: 1056673 · Received March 28, 2008

Report

Report Number
9611451-2008-00134
Event Type
Malfunction
Date Received
March 28, 2008
Date of Event
February 6, 2008
Report Date
February 6, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
MNT
PMA / PMN Number
K060044
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

METHOD: THE INVESTIGATION WAS CONDUCTED BASED ON THE RETURNED MASKS, EVENT DESCRIPTION AND PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS. RESULTS: THE SEALS WERE HELD TO THE MASK BASES BY FRICTION. THE FRICTION APPLIED WAS NOT SUFFICIENT ENOUGH TO HOLD THE SEALS TO THE BASES OF THE MASKS. LOT NUMBER 071015: OUR RECORDS INDICATE THAT TWO COMPLAINTS OF THIS NATURE HAVE BEEN RECEIVED FOR THE GIVEN LOT NUMBER. LOT NUMBER 071005: OUR RECORDS INDICATE THAT THIS IS THE ONLY COMPLAINT OF THIS NATURE RECEIVED FOR THE GIVEN LOT NUMBER. CONCLUSION: WE HAVE AN OPEN CAPA PROJECT TO ADDRESS THIS ISSUE AND TO IMPLEMENT POTENTIAL DESIGN SOLUTIONS WE HAVE DEVELOPED TO PREVENT SEPARATION ISSUES OCCURRING IN THE FUTURE. WE ARE CURRENTLY IMPLEMENTING AN ADDED SILICONE ADHESIVE STEP (THE APPLICATION OF AN ADHESIVE BETWEEN THE SILICONE FACIAL SEAL AND THE PLASTIC MASK BASE) IN OUR PRODUCTION PROCESS. ADD'L LOT NUMBER: 071005.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN COLUMBUS, OHIO REPORTED TO OUR DISTRIBUTOR THAT 3 SEALS OF RT040L FULL FACE MASKS HAD SEPARATED FROM THE MASK BASES PRIOR TO THE OPENING OF THE PACKAGE. NO PT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FULL FACE MASK MNT FISHER & PAYKEL HEALTHCARE, LTD. RT040L 071015

Patients

Seq Age Sex Outcome Treatment
1