FULL FACE MASK
Report
- Report Number
- 9611451-2008-00135
- Event Type
- Malfunction
- Date Received
- June 4, 2008
- Date of Event
- February 6, 2008
- Report Date
- February 6, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- MNT
- PMA / PMN Number
- K060044
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
ADD'L LOT NUMBERS ARE 07116 AND 070904. ADD'L MFG DATES ARE 11/6/2007 AND 9/4/2007. THE INVESTIGATION WAS CONDUCTED BASED ON THE RETURNED MASKS, EVENT DESCRIPTION AND PREVIOUS INVESTIGATIONS ON SIMILAR COMPLAINTS. RESULTS: THE SEALS WERE HELD TO THE MASK BASES BY FRICTION. THE FRICTION APPLIED WAS NOT SUFFICIENT ENOUGH TO HOLD THE SEALS TO THE BASES OF THE MASKS. LOT NUMBER 071108: OUR RECORDS INDICATE THAT THIS IS THE ONLY COMPLAINT OF THIS NATURE RECEIVED FOR THE GIVEN LOT NUMBER. LOT NUMBER 071106: OUR RECORDS INDICATE THAT THIS IS THE ONLY COMPLAINT OF THIS NATURE RECEIVED FOR THE GIVEN LOT. LOT NUMBER 070904:OUR RECORDS INDICATE THAT THIS IS THE ONLY COMPLAINT OF THIS NATURE RECEIVED FOR THE GIVEN LOT NUMBER. CONCLUSION: WE HAVE AN OPEN CAPA PROJECT TO ADDRESS THIS ISSUE AND TO IMPLEMENT POTENTIAL DESIGN SOLUTIONS WE HAVE DEVELOPED TO PREVENT SEPARATION ISSUES OCCURRING IN THE FUTURE. WE ARE CURRENTLY IMPLEMENTING AN ADDED SILICONE ADHESIVE STEP (THE APPLICATION OF AN ADHESIVE BETWEEN THE SILICONE FACIAL SEAL AND THE PLASTIC MASK BASE) IN OUR PRODUCTION PROCESS.
A HEALTHCARE FACILITY REPORTED TO OUR DISTRIBUTOR THAT THE SEALS OF THREE RT040S FULL FACE MASKS WERE DETACHED FROM THE MASK BASES PRIOR TO THE OPENING OF THE PACKAGE. NO PT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FULL FACE MASK | MNT | FISHER & PAYKEL HEALTHCARE, LTD. | RT040S | 071108 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |