GLUCOCARD EXPRESSION BLOOD GLUCOSE TEST SYSTEM

MDR report key: 10564824 · Received September 21, 2020

Report

Report Number
9616936-2020-00002
Event Type
Injury
Date Received
September 21, 2020
Date of Event
August 3, 2020
Report Date
August 4, 2020
Manufacturer
APEX BIOTECHNOLOGY CORP.
Product Code
NBW
UDI-DI
00015482570057
PMA / PMN Number
K113098
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FROM IMPORTER (ARKRAY USA INC,, HEREAFTER : ARKRAY). ARKRAY NOTIFIED THE MANUFACTURER (APEX BIOTECHNOLOGY CORP., HEREAFTER: APEXBIO) ON 25 AUG 2020 THAT THEY HAVE RECEIVED A CUSTOMER COMPLAINT WHICH WAS A MDR. ARKRAY RECEIVED THE COMPLAINED METER AND TEST STRIP (1 PIECE) AND PERFORMED TESTS. RESULTS WERE PASS. INVESTIGATION FROM MANUFACTURER (APEXBIO): APEXBIO PERFORMED FUNCTIONAL TEST TO THE RETURNED METER. ALL TESTS WERE PASS. APEXBIO PERFORMED VENOUS BLOOD TEST AND CONTROL SOLUTION TEST FOR RETURNED METER AND RETAIN STRIPS OF COMPLAINED LOT OF TEST STRIP. RESULTS WERE PASS. APEXBIO PERFORMED ADDITIONAL MOISTURE TEST TO THE DESICCANT IN STRIP BOTTLE. RESULTS WERE ACCEPTABLE.

Description of Event or Problem · 1

CUSTOMER'S WIFE WAS CALLING TO REPORT THAT THE METER WAS READING TOO HIGH. READING IS TOO HIGH BASED ON SYMPTOMS DISPLAYED AND CALLER STATED IF READING WAS ACCURATE, CUSTOMER WOULDN'T HAVE BEEN DISPLAYING SYMPTOMS. PATIENT WAS EATING LUNCH AND TOOK INSULIN ON HIS PUMP. THEN HE CALLED TO HIS WIFE AND ASKED FOR HELP AS HE WAS EXPERIENCING SEVERE HYPOGLYCEMIC SYMPTOMS AND HAVING A HARD TIME TALKING. HE WAS THEN CHECKED ON THE EXPRESSION METER AND THE READING WAS 50. HE WAS THEN CHECKED ON HIS BAYER METER AND THE READING WAS 31. HE WAS GIVEN JUICE AND GLUCOSE GEL AFTER 15 MINUTES HE WAS STILL VERY LOW, AND THE PARAMEDICS WERE CALLED. HE WAS GIVEN MORE GLUCOSE GEL FROM THEM AND IT TOOK TWO HOURS TO GET HIS BLOOD SUGAR UP. HE WAS NOT TAKEN TO THE HOSPITAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1028835 GLUCOCARD EXPRESSION BLOOD GLUCOSE TEST SYSTEM GLUCOCARD EXPRESSION BLOOD GLUCOSE TEST SYSTEM NBW APEX BIOTECHNOLOGY CORP. 571100 B4S1911009 00015482570057

Patients

Seq Age Sex Outcome Treatment
1 70 YR Life Threatening| R