FDA Adverse Event Malfunction Summary report: N

NI

MDR report key: 10564814 · Received September 21, 2020

Report

Report Number
8030229-2020-00552
Event Type
Malfunction
Date Received
September 21, 2020
Date of Event
August 22, 2020
Report Date
January 25, 2023
Manufacturer
NIHON KOHDEN CORPORATION
Product Code
MHX
PMA / PMN Number
NI
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DETAILS OF COMPLAINT: THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE CENTRAL NURSE'S STATIONS (CNS) AND THE REMOTE NETWORK STATIONS (RNS) WENT DOWN DUE TO A POWER ISSUE WITH THE SERVERS. THEY CORRECTED THE POWER ISSUE, AND THE CNSS CAME BACK UP NORMALLY, BUT THE RNSS WERE IN COMM LOSS. NIHON KOHDEN COMPUTER INFORMATION TECHNOLOGY SYSTEMS SUPPORT (CITS) RESTARTED THE UNIFIED GATEWAY (UG) SERVICE, WHICH RESOLVED THE ISSUE WITH THE RNSS. NO PATIENT HARM OR INJURY WAS REPORTED. INVESTIGATION SUMMARY: THE ISSUE WAS RESOLVED WHEN CITS REMOTED INTO THE SYSTEM AND RESET THE UG SERVICE. THE ROOT CAUSE IS DETERMINED TO BE THE FACILITY'S NETWORK ENVIRONMENT. AN UNGRACEFUL EXIT DUE TO A POWER OUTAGE RESULTED IN NETWORK SETTINGS DEFAULTING TO THE FACTORY SETTINGS, WHICH WERE RESOLVED WITH THE UG SERVICE REBOOT.

Description of Event or Problem · 0

THE BIOMEDICAL ENGINEER REPORTED THAT THEY HAD A POWER ISSUE WITH THE SERVERS AND THE CENTRAL NURSE'S STATIONS (CNS) AND THE REMOTE NETWORK STATIONS (RNS) WENT DOWN. THEY CORRECTED THE ISSUE WITH THE POWER AND THE CNSS CAME BACK UP NORMALLY, BUT THE RNSS ALL SHOWED AS COMMUNICATION LOSS. NIHON KOHDEN COMPUTER INFORMATION TECHNOLOGY SYSTEMS SUPPORT (CITS) RESTARTED THE UNIFIED GATEWAY SERVICE, AND THE RNSS WERE NOW ABLE TO MONITOR PATIENTS. NO PATIENT HARM WAS REPORTED.

Additional Manufacturer Narrative · 1

THE BIOMEDICAL ENGINEER REPORTED THAT THEY HAD A POWER ISSUE WITH THE SERVERS, THE CENTRAL NURSE'S STATIONS (CNS), AND THE REMOTE NETWORK STATIONS (RNS) WENT DOWN. THEY CORRECTED THE ISSUE WITH THE POWER, AND THE CNSS CAME BACK UP NORMALLY, BUT THE RNSS ALL SHOWED AS COMMUNICATION LOSS. NIHON KOHDEN COMPUTER INFORMATION TECHNOLOGY SYSTEMS SUPPORT (CITS) RESTARTED THE UNIFIED GATEWAY SERVICE, AND THE RNSS WERE NOW ABLE TO MONITOR PATIENTS. NO PATIENT HARM WAS REPORTED. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. CONCOMITANT MEDICAL DEVICE: THE FOLLOWING DEVICE(S) WERE BEING USED IN CONJUNCTION WITH THE CNS. REMOTE NETWORK STATION: MODEL: RNS-6803, SN: (B)(4).

Description of Event or Problem · 1

THE BIOMEDICAL ENGINEER REPORTED THAT THEY HAD A POWER ISSUE WITH THE SERVERS, THE CENTRAL NURSE'S STATIONS (CNS), AND THE REMOTE NETWORK STATIONS (RNS) WENT DOWN. THEY CORRECTED THE ISSUE WITH THE POWER, AND THE CNSS CAME BACK UP NORMALLY, BUT THE RNSS ALL SHOWED AS COMMUNICATION LOSS. NIHON KOHDEN COMPUTER INFORMATION TECHNOLOGY SYSTEMS SUPPORT (CITS) RESTARTED THE UNIFIED GATEWAY SERVICE, AND THE RNSS WERE NOW ABLE TO MONITOR PATIENTS. NO PATIENT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1028732 NI CENTRAL MONITOR STATION MHX NIHON KOHDEN CORPORATION NI NA

Patients

Seq Age Sex Outcome Treatment
1 Unknown RNS-6803 SN (B)(6)| RNS-6803 SN (B)(6)