FDA Adverse Event Malfunction Summary report: N

RI ROBOTIC DRILL ATTACHMENT

MDR report key: 10564584 · Received September 21, 2020

Report

Report Number
3010266064-2020-01784
Event Type
Malfunction
Date Received
September 21, 2020
Date of Event
August 26, 2020
Report Date
April 13, 2021
Manufacturer
BLUE BELT TECHNOLOGIES
Product Code
OLO
UDI-DI
00885556757345
PMA / PMN Number
K193120
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DRILL ATTACHMENT INTENDED FOR USE IN TREATMENT WAS RETURNED. NOTHING WAS IDENTIFIED VISUALLY THAT CONTRIBUTED TO THE REPORTED PROBLEM. A FUNCTIONAL EVALUATION WAS PERFORMED. A KPC TEST WAS PERFORMED AND THE TEST PASSED. THE REPORTED PROBLEM WAS NOT CONFIRMED. THE DRILL ATTACHMENT FUNCTIONED AS EXPECTED WITHOUT ANY RESISTANCE ISSUES. AN ENGINEERING EVALUATION WAS PERFORMED. THERE ARE NO ISSUES. PASSED THE KPC TEST WITH THE SAME VALUES AS A NORMAL LONG ATTACHMENT. A REVIEW OF MANUFACTURING AND SERVICE RECORDS INDICATE THE DEVICE MET ALL SPECIFICATIONS UPON RELEASE INTO DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOR SIMILAR REPORTED/CONFIRMED COMPLAINTS HAS IDENTIFIED PRIOR EVENTS. ALTHOUGH THE REPORTED PROBLEM WAS NOT CONFIRMED, NO REASONABLE CONTRIBUTING FACTORS COULD BE IDENTIFIED BASED ON THE RECEIVED COMPLAINT INFORMATION AND INVESTIGATION RESULTS. BASED ON THE INVESTIGATION, NO FURTHER CONTAINMENT OR CORRECTIVE ACTION IS RECOMMENDED OR REQUIRED AT THIS TIME. INTERNAL COMPLAINT REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER REPLACING BOTH THE CAMERA AND THE CORI CONSOLE, CORI SAWBONE LAB/DEMO WAS PLANNED AND THE LONG ATTACHMENTS WERE CHECKED RESPECTIVELY WITH A BUR. THE BUR WAS INSERTED IN THE LONG ATTACHMENT TO SEE THE RESISTANCE PRESENTED TO THE BUR, AND THREE LONG ATTACHMENTS PRESENTED RESISTANCE. THE BEARINGS WITHIN THE ATTACHMENTS COULD HAVE AN EXCESS OF EPOXY LIKELY CREATING FRICTION WHEN SPINNING THE BUR WHICH CAN ALSO CREATE HEAT. THIS EXPLAINS WHY SOME OF THE BUR ATTACHMENTS ARE MUCH HOTTER THAN OTHERS WHILE BURRING. NO PATIENT WAS INVOLVED. NO OTHER COMPLICATIONS WERE REPORTED.