PISCES QUAD PLUS
Report
- Report Number
- 2182207-2008-02970
- Event Type
- Injury
- Date Received
- June 4, 2008
- Report Date
- May 5, 2008
- Manufacturer
- MEDTRONIC NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- K923567
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
.
JOURNAL REFERENCE: TRENTMAN TL, DODICK DW, ZIMMERMAN RS, BIRCH BD. PERCUTANEOUS OCCIPITAL STIMULATOR LEAD TIP EROSION: REPORT OF 2 CASES. PAIN PHYS 2008; 11(2):253-256. OCCIPITAL NERVE STIMULATION IS AN EMERGING TREATMENT MODALITY FOR REFRACTORY HEADACHE DISORDERS. WE DESCRIBE THE FIRST 2 REPORTED CASES OF PERCUTANEOUS OCCIPITAL STIMULATOR LEAD TIP EROSION. IN BOTH CASES, THE LEAD EROSION OCCURRED MANY MONTHS AFTER IMPLANTATION. ONE PT LOST A SIGNIFICANT AMOUNT OF WEIGHT BETWEEN THE TIME OF IMPLANT AND LEAD EROSION, WHILE THE OTHER PT HAD NO OBVIOUS RISK FACTORS. BOTH PTS RETURNED TO EXCELLENT HEADACHE CONTROL. REPORTABLE EVENT: A WOMAN WITH LEFT OCCIPITAL AND PARIETAL REGION HEADACHES. SHE UNDER WENT AN ONS TRAIL WITH UNILATERAL PERCUTANEOUS LEAD. DURING TRIAL SHE HAD COMPLETE PAIN RELIEF AND WAS IMPLANTED WITH A PERMANENT SYSTEM 1 MONTH LATER. SIX MONTHS LATER, SHE COMPLAINED OF SORENESS AND A PALPABLE LEAD TIP IN THE RETROMASTOID REGION. THERE WAS NO INFECTION AND A PALPABLE LEAD TIP IN THE RETROMASTOID REGION. THERE WAS NO INFECTION AND OBSERVATION WAS ADVISED. IN 16 MONTHS LATER, SHE PRESENTED WITH 2 WEEKS OF DISCOMFORT AND ACUTE LEAD TIP EROSION WITH LOCALIZED INFECTION. THE LEAD WAS REMOVED, SHE WAS TREATED WITH ANTIBIOTICS AND 1 MONTH LATER, THE LEAD WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PISCES QUAD PLUS | GZB | MEDTRONIC NEUROMODULATION | 3888 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Required Intervention | PROGRAMMER MODEL UNK N=1| IMPLANTABLE PULSE GENERATOR MODEL UNK N=1| LEAD MODEL 3888 N=1 |