FDA Adverse Event Malfunction Summary report: N

HOMECHOICE CYCLER-REFURBISHED

MDR report key: 1056078 · Received June 3, 2008

Report

Report Number
1423500-2008-00439
Event Type
Malfunction
Date Received
June 3, 2008
Date of Event
May 13, 2008
Report Date
May 13, 2008
Manufacturer
BAXTER HEALTHCARE
Product Code
FKX
PMA / PMN Number
K012988
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE CYCLER'S INITIAL DRAIN ALARM SETPOINT WAS PROGRAMMED TOO LOW AT 0 MLS, RESULTING IN AN INCOMPLETE INITIAL DRAIN FOLLOWED BY A FULL FILL. AS A RESULT OF THIS INCIDENT, THE INITIAL DRAIN ALARM SETPOINT WAS INCREASED FROM 0MLS TO 1500MLS. ACCORDING TO THE HP'S HEALTHCARE PROFESSIONAL, REPROGRAMMING OF THE CYCLER HAS RESOLVED THIS INCIDENT AND THE HP CONTINUES TO USE THE HOMECHOICE CYCLER WITHOUT FURTHER REPORTED DIFFICULTY. DEVICE REPROGRAMMING HAS RESOLVED THIS INCIDENT, THEREFORE, AN EVALUATION OF THE DEVICE WAS NOT REQUIRED.

Description of Event or Problem · 1

THE HOME PATIENT (HP) CONTACTED A TECHNICAL SERVICE REPRESENTATIVE (TSR) AND REPORTED ABDOMINAL DISTENTION, AS IF HE WERE OVERFILLED WITH FLUID, DURING THERAPY USING THE HOMECHOICE AUTOMATED PERITONEAL DIALYSIS (APD) SYSTEM. THE INCIDENT WAS REVIEWED OVER THE PHONE WITH THE TSR, WHICH REVEALED THE HP DRAINED 362MLS DURING THE INITIAL DRAIN WHEN THE CYCLER ADVANCED INTO DAY FILL 1. THE HP RECEIVED THE DAY FILL VOLUME OF 2000MLS AND EXPERIENCED ABDOMINAL DISTENTION. DURING THE DAY DRAIN THE HP REMOVED 3300 MLS OF FLUID, WHICH WAS 1100MLS GREATER THAN THE PROGRAMMED FILL VOLUME OF 2000 MLS. REVIEW OF THE CYCLER'S PROGRAMMING REVEALED THE HP WAS PERFORMING HI-DOSE CONTINUOUS CYCLING PERITONEAL DIALYSIS (CCPD), DAY FILL VOLUME = 2000MLS, NIGHT FILL VOLUME = 2000MLS, LAST FILL VOLUME = 2000MLS, INITIAL DRAIN ALARM SETPOINT = 0MLS. THE TSR REVIEWED HOMECHOICE DRAIN LOGIC WITH THE HP AND CONTACTED THE HP'S DIALYSIS CENTER NURSE (RN). AS A RESULT OF THIS INCIDENT, THE TSR VERBALLY ASSISTED THE HP TO INCREASE THE INITIAL DRAIN ALARM SETPOINT FROM 0 MLS TO 1500 MLS. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE CYCLER-REFURBISHED 78FKX FKX BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1