FDA Adverse Event Malfunction Summary report: N

BIOGX SARS-COV-2 OPEN SYSTEM REAGENTS FOR BD MAX SYSTEM

MDR report key: 10560340 · Received September 21, 2020

Report

Report Number
1119779-2020-00377
Event Type
Malfunction
Date Received
September 21, 2020
Date of Event
September 1, 2020
Report Date
November 18, 2020
Manufacturer
BECTON, DICKINSON & CO.
Product Code
QJR
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY THE COMPLAINT INVESTIGATION FOR DISCREPANT RESULTS USING THE BIOGX SARS-COV-2 OSR FOR BD MAX SYSTEM (REF 444213) (UNKNOWN LOT #) ALONG WITH THE KIT BD MAX EXK TNA 3 PRODUCT LOT 0182268 WAS PERFORMED BY BD. SINCE NO KIT LOT # WAS PROVIDED BY THE CUSTOMER, BIOGX COULD NOT PERFORM THE MANUFACTURING REVIEW ON THE BIOGX SARS-COV-2 OSR SPECIFIC LOT. THE INVESTIGATION WAS PERFORMED BY THE REVIEW OF MANUFACTURING RECORDS ON THE BD MAX EXK TNA 3 LOT 0182268, THE REVIEW OF CUSTOMER DATA AND THE VERIFICATION OF COMPLAINTS HISTORY. REVIEW OF THE MANUFACTURING RECORDS OF THE BD MAX EXK TNA 3 INDICATED THAT THE LOT WAS MANUFACTURED ACCORDING TO SPECIFICATIONS. CUSTOMER REPORTED HIGH POSITIVE RESULTS AND PROVIDED DATABASE FROM INSTRUMENT CT1735 FOR ANALYSIS. THE RUN #103 NAMED IN THE COMPLAINT TEXT CONTAINED MOSTLY POSITIVE RESULTS, BUT CUSTOMER DID NOT MENTION WHICH SAMPLE WAS DISCREPANT. ANALYSIS OF THE DATABASE SHOW THAT SOME ENVIRONMENTAL MONITORING WAS DONE AND MANY OF THE SITES SAMPLE TESTED WERE POSITIVE. THE RUN #116 SHOWS THAT ALL SURFACE SWABBED GAVE POSITIVE RESULTS, SOMETIMES ONLY FOR N1 AND OTHER TIMES FOR BOTH N1 AND N2. ANALYSIS OF THE CURVES REVEALED HIGH FLUORESCENCE VALUES OF EACH SAMPLE WITH TRUE AMPLIFICATION. MOREOVER, AN ENVIRONMENTAL SAMPLE TAKEN ON QIAQUANT COUNTERTOP OBTAINED CT VALUES AROUND 22 FOR BOTH TARGETS, WHICH REPRESENTS A HIGH LOAD. AN ENVIRONMENTAL CONTAMINATION IS STRONGLY SUSPECTED, AND THE QIAQUANT COUNTERTOP COULD BE THE SOURCE. THERE IS NO INDICATION OF AN INCREASE IN COMPLAINTS FOR DISCREPANT RESULTS COMPLAINTS FOR THE BIOGX SARS-COV-2 OSR PRODUCT. THE ROOT CAUSE WAS NOT IDENTIFIED. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. BD & BIOGX PRODUCT MANUFACTURERS DID NOT INITIATE A CORRECTIVE AND PREVENTIVE ACTION (CAPA). EUA# (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BIOGX SARS-COV-2 OPEN SYSTEM REAGENTS FOR BD MAX¿ SYSTEM, THERE WERE HIGH FALSE POSITIVE RESULTS OBTAINED BY THE LABORATORY PERSONNEL. THERE IS NO INDICATION THAT ERRONEOUS RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION FROM THE CUSTOMER. EUA#(B)(4).

Additional Manufacturer Narrative · 1

EUA# (B)(4). MEDICAL DEVICE LOT #: 0182268 WAS REPORTED, HOWEVER, THIS IS NOT A LOT# MANUFACTURED FOR THIS PRODUCT. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE BIOGX SARS-COV-2 OPEN SYSTEM REAGENTS FOR BD MAX¿ SYSTEM, THERE WERE HIGH FALSE POSITIVE RESULTS OBTAINED BY THE LABORATORY PERSONNEL. THERE IS NO INDICATION THAT ERRONEOUS RESULTS WERE REPORTED OUT, AND THERE WAS NO REPORT OF PATIENT IMPACT. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION FROM THE CUSTOMER. EUA# (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1026887 BIOGX SARS-COV-2 OPEN SYSTEM REAGENTS FOR BD MAX SYSTEM SARS-COV-2 REAGENT KIT QJR BECTON, DICKINSON & CO. SEE H.10

Patients

Seq Age Sex Outcome Treatment
1 Other