8015 ALARIS SYSTEM PC UNIT
Report
- Report Number
- 2016493-2020-09702
- Event Type
- Malfunction
- Date Received
- September 21, 2020
- Report Date
- January 23, 2018
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- PMA / PMN Number
- K091308
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND ANY SUBSEQUENT REPAIRS HAVE BEEN INVESTIGATED THROUGH THE NORMAL SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS USED INFORMATION, WAS REVIEWED FOR THE SAP FILE AND TRACK WISE FILE, AND FOUND TO BE RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE SOURCE DEVICE SERVICE HISTORY RECORD WAS PERFORMED BEGINNING FROM THE DATE OF MANUFACTURE TO THE PRESENT DATE AND INDICATED THAT THIS DEVICE HAS NOT BEEN PREVIOUSLY RETURNED FOR SERVICE. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE SOURCE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE AND PRODUCT WAS RETURNED FOR THE SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. CAPA REFERENCE: (B)(4). THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(6) 2018 12:17:02 (B)(6). WAR: PO: (B)(6). SHIPPING & BILLING ADDRESSES CONFIRMED. NPI. (B)(6) 2018 20:11:19 (B)(6). REPLACED C64; C270 ON LOGIC BOARD. (B)(6) 2018 20:13:03 (B)(6). VERIFIED SOLDER C64 AND C270 ON LOGIC BOARD. (B)(6) 2018 08:09:13 (B)(6). (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1027763 | 8015 ALARIS SYSTEM PC UNIT | PUMP, INFUSION | FRN | CAREFUSION SD | 8015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |