FDA Adverse Event Malfunction Summary report: N

STAR

MDR report key: 10557300 · Received September 18, 2020

Report

Report Number
3006695864-2020-00438
Event Type
Malfunction
Date Received
September 18, 2020
Date of Event
August 27, 2020
Report Date
September 18, 2020
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC.
Product Code
LZS
UDI-DI
00(01)(21)3485
PMA / PMN Number
P930016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SYSTEM WAS EVALUATED BY A FIELD SERVICE ENGINEER. THE FIELD SERVICE REPLACED SYSTEM SAM-C CARD, REPLACED BOTH CARD READERS. A FIELD SERVICE CHECKLIST WAS PERFORMED. THE UNIT COMPLIED WITH ALL FACTORY SETTINGS. A RECORD REVIEW WAS PERFORMED. A PRODUCT DEFICIENCY REVIEW WAS PERFORMED AND THERE IS NO PRODUCT DEFICIENCY IDENTIFIED. A DOCUMENT, SERVICE HISTORY, AND TRENDING WAS REVIEWED. THERE IS NOT A RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. A LABELING REVIEW WAS CONDUCTED; THE OPERATOR MANUAL FOR THE SYSTEM WAS REVIEWED AND FOUND TO INCLUDE ADEQUATE INSTRUCTIONS FOR USE, WARNINGS AND OPERATIONAL ERRORS. THE REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR STAR LASER SYSTEM SHOWED THAT THERE WERE NO ISSUES OR NON-CONFORMITIES. THE SYSTEM AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. MANUFACTURING HAS BEEN RULED OUT AS A POTENTIAL CAUSE FOR THE REPORTED ISSUE. BASED ON THE INVESTIGATION RESULTS, NO CORRECTIVE ACTION HAS BEEN ISSUED. BASED ON THE INVESTIGATION RESULTS THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. JOHNSON & JOHNSON SURGICAL VISION WILL CONTINUE TO MONITOR THIS TYPE OF COMPLAINTS. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A LASER VISION CORRECTION PATIENT CHAIR WAS MOVING ON ITS OWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1016336 STAR EXCIMER LASER LZS JOHNSON & JOHNSON SURGICAL VISION, INC. 0030-1479 00(01)(21)3485

Patients

Seq Age Sex Outcome Treatment
1