EDWARDS SAPIEN XT TRANSCATHETER HEART VALVE
Report
- Report Number
- 2015691-2020-13645
- Event Type
- Injury
- Date Received
- September 18, 2020
- Date of Event
- August 25, 2020
- Report Date
- August 26, 2020
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- NPT
- PMA / PMN Number
- P130009
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- PHYSICIAN
Narratives
UDI: UNKNOWN. PER THE INSTRUCTIONS FOR USE (IFU), PARAVALVULAR LEAK (PVL) IS A POTENTIAL ADVERSE EVENT ASSOCIATED WITH BIOPROSTHETIC HEART VALVES. PARAVALVULAR LEAK REFERS TO BLOOD FLOWING THROUGH A CHANNEL BETWEEN THE STRUCTURE OF THE IMPLANTED VALVE AND THE CARDIAC TISSUE, AS A RESULT OF A LACK OF APPROPRIATE SEALING OF THE VALVE TO THE TARGET SITE. SOME PVL IS NOT UNCOMMON POST DEPLOYMENT. MANY CASES ARE MILD TO MODERATE, AND EITHER RESOLVE OVER TIME OR DO NOT CAUSE SYMPTOMS. OTHERS MAY BE MORE CLINICALLY SIGNIFICANT AND REQUIRE INTERVENTION. THE MECHANISM BEHIND WORSENING OR LATE PVL IS NOT WELL UNDERSTOOD BUT MAY BE RELATED TO CARDIAC REMODELING. IN THIS CASE, THERE WAS NO ALLEGATION OR INDICATION A PRODUCT DEFICIENCY CONTRIBUTED TO THIS ADVERSE EVENT. THE EXACT CAUSE OF THE REPORTED EVENT WAS UNABLE TO BE DETERMINED, HOWEVER MAY BE RELATED TO PATIENT AND/OR PROCEDURAL FACTORS NOT PROVIDED. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED.
AS REPORTED BY A FIELD CLINICAL SPECIALIST APPROXIMATELY 6 YEARS POST IMPLANT OF A 23 MM SAPIEN 3 XT VALVE IN THE AORTIC POSITION VIA TRANSFEMORAL APPROACH SEVERE PVL WAS OBSERVED. A VALVE IN VALVE PROCEDURE WAS PERFORMED WITH A 20 MM SAPIEN 3 VALVE. POST PROCEDURE THE PATIENT IS DOING WELL IN STABLE CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1016837 | EDWARDS SAPIEN XT TRANSCATHETER HEART VALVE | AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED | NPT | EDWARDS LIFESCIENCES | 9300TFX23 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Required Intervention |