FDA Adverse Event Malfunction Summary report: N

LOCALIZER

MDR report key: 10556600 · Received September 18, 2020

Report

Report Number
3013649990-2020-00003
Event Type
Malfunction
Date Received
September 18, 2020
Date of Event
August 18, 2020
Report Date
August 18, 2020
Manufacturer
HEALTH BEACONS, INC.
Product Code
NEU
PMA / PMN Number
K193189
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

JUSTIFICATION FOR THE SUBMISSION OF THE MDR PAST THE REQUIRED 30 DAYS: DAY 30: ON (B)(6) 2020, EMDR WAS PACKAGED FOR SUBMISSION. WHEN CONNECTED TO WEBTRADER TO SUBMIT REPORT, THE SYSTEM REJECTED THE REPORT SUBMISSION DUE TO AN EXPIRED CERTIFICATE. IT WAS NOT POSSIBLE TO RENEW THE CERTIFICATE ON 17-SEP-2020. DAY 31: ON (B)(6) 2020, AN ALTERNATE WAY TO SUBMIT THE REPORT THROUGH PARENT COMPANY (HOLOGIC INC.) WAS DECIDED UPON. REPORT IS BEING SUBMITTED ONE DAY LATE.

Description of Event or Problem · 1

ON (B)(6) 2020, HEALTH BEACONS BECAME AWARE OF AN INCIDENT AT (B)(6). PATIENT WAS HAVING A HEALTH BEACONS TAG PLACED WITH A 10CM TAG APPLICATOR. ACCORDING TO THE RN, THE APPLICATOR WAS DIFFICULT TO PUSH THROUGH THE BREAST, AS THE PATIENT'S BREAST WAS VERY DENSE. AFTER EVALUATING PLACEMENT OF THE NEEDLE WITH IMAGES, THE TAG DEPLOYED AND AT SOME POINT, THE TAG APPEARS ON IMAGING TO HAVE SPLIT IN TWO PIECES. THE TAG DID NOT GIVE A SIGNAL AFTER IMPLANTATION, BUT THE SIGNAL AND TAG ID# HAD BEEN CONFIRMED DURING INTIAL SCANNING OF THE PACKAGING PRIOR TO IMPLANTATION. PATIENT BECAME VASOVAGAL DUE TO THE FACT SHE WAS SITTING UP FOR THE UPRIGHT STEREOTACTIC PLACEMENT OF THE TAG. THE PHYSICIAN DETERMINED THERE WAS ENOUGH EVIDENCE FOR ORIENTATION TO REMOVE THE AREA OF INTEREST WITHOUT DIFFICULTY, AND THE PATIENT'S SURGERY WAS SCHEDULED FOR THE FOLLOWING WEEK AS ORIGINALLY INTENDED. DATE OF THE SURGERY WAS NOT MOVED. ON (B)(6) 2020, HEALTH BEACONS SPOKE WITH THE CLINICAL SPECIALIST, WHO CONFIRMED THE REMOVAL OF THE TISSUE, AND THE TAG WERE SUCCESSFUL, AND THAT THE RECOVERED TAG HAD BEEN SENT WITH THE TISSUE TO PATHOLOGY. CLINICAL SPECIALIST EXPRESSED CONFIDENCE THAT THE CAUSE OF THE TAG BREAKAGE WAS DELIVERING THE TAG AGAINST RESISTANCE. THEY HAVE ENCOUNTERED THIS BEFORE, AND IN ANOTHER CASE HAD DETERMINED TO USE AN INTRODUCER TO AID PLACEMENT. ON 11-SEP-2020, THE FIELD REPRESENTATIVE SHARED IMAGES OF THE EXPLANTED TAG, WHICH CLEARLY SHOW THE TAG BROKEN IN TWO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1016676 LOCALIZER TAG APPLICATOR NEU HEALTH BEACONS, INC. HB200-10 45258

Patients

Seq Age Sex Outcome Treatment
1