GALILEO
Report
- Report Number
- 1034569-2008-00173
- Event Type
- Malfunction
- Date Received
- June 5, 2008
- Date of Event
- May 12, 2008
- Report Date
- May 30, 2008
- Manufacturer
- IMMUCOR, INC.
- Product Code
- KSZ
- PMA / PMN Number
- BK040013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
FWD_ABORH TESTING WAS PERFORMED ON AN IN-HOUSE GALILEO WITH IN-HOUSE DONOR SAMPLES OF VARIOUS ABO AND RH TYPES USING RETENTION ANTI-A, LOT 101679, ANTI-B SERIES 3, LOT 203234, ANTI-D SERIES 4, LOT 504697, AND MONOCLONAL CONTROL, LOT 492028. ALL IN-HOUSE DONOR SAMPLES TYPED AS EXPECTED WITH NO DISCREPANCIES. TESTING PERFORMED WITH RETENTION ANTI-A, LOT 101673, 101674 WITH IN HOUSE DONOR SAMPLES REVEALED TYPE A RED CELL STICKING IN THE ANTI-A WELLS. TO PREVENT THE STICKING, IN HOUSE STUDIES FOUND THAT PIPETTING THE REAGENTS INTO THE WELLS BEFORE THE SAMPLE RED CELLS GREATLY REDUCED THE STICKING. IN HOUSE COMPARISON TESTING OF THE NEW REAGENT-FIRST PIPETTING ORDER TO THE EXISTING SAMPLE-FIRST PIPETTING ORDER SHOWED NO UNEXPECTED NEGATIVE ANTI-A REACTIONS WITH THE NEW PIPETTING ORDER. SOME UNEXPECTED NEGATIVE REACTIONS WERE OBSERVED DURING TESTING WITH THE CURRENT ASSAY. CUSTOMERS WERE NOTIFIED OF THE MODIFICATION TO THE ASSAY ON 9/20/07.THIS FACILITY DID HAVE THE MODIFIED ABO ASSAY SOFTWARE INSTALLED. ACCORDING TO THE GALILEO OPERATOR MANUAL,FORWARD ONLY ABO-RH TESTING HAS A HIGHER RISK OF MISTYPE DUE TO THE ABSENCE OF THE REVERSE TYPE RESULTS. HAZARDOUS MISTYPES MAY OCCUR, SUCH AS A SAMPLE BEING INTERPRETED AS A GROUP AB, OR AN RH (D) NEGATIVE SAMPLE BEING INTERPRETED AS RH(D) POSITIVE. FOR THIS REASON, ABO-RH RESULTS SHOULD ALWAYS BE COMPARED TO THE PATIENT OR DONOR'S HISTORY.
CUSTOMER REPORTED THAT FIVE DONOR SPECIMENS WERE TESTED WITH THE GALILEO (FWD_ABORH ASSAY) AND FOUND TO BE GROUP O; HISTORICAL TYPING OF THESE DONORS REVEALED A TYPE OF GROUP A.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GALILEO | AUTOMATED BLOOD BANK ANALYZER | KSZ | IMMUCOR, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |