BD ULTRA FINE PEN NEEDLE
Report
- Report Number
- 9616656-2020-00930
- Event Type
- Malfunction
- Date Received
- September 17, 2020
- Date of Event
- September 8, 2020
- Report Date
- September 11, 2020
- Manufacturer
- BECTON DICKINSON AND CO.
- Product Code
- FMI
- UDI-DI
- 10885403928659
- PMA / PMN Number
- K162516
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- 003
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. (B)(4). INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST. RELATED COMPLAINT FOR NEEDLE CLOG ON LOT # 0022825. A REVIEW OF RISK MANAGEMENT DOCUMENT (B)(4), INDICATES THAT THE POTENTIAL RISK OF THIS SPECIFIC REPORTED INCIDENT (PEN NEEDLE, NEEDLE CLOG) WAS CAPTURED AND ADDRESSED. INVESTIGATION CONCLUSION: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED. ROOT CAUSE DESCRIPTION: ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. RATIONALE: BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.
IT WAS REPORTED THAT THE BD ULTRA FINE¿ PEN NEEDLE WAS CLOGGED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CONSUMER REPORTED FINDING ABOUT 50 % FROM THIS BOX OF PEN NEEDLES CLOG DURING INJECTION. INFORMED OF PROPER PLACEMENT AND TO COMPLETE FLOW CHECK WITH ALL INJECTIONS. SHE COMPLETES FLOW CHECK ONLY WITH SOME INJECTIONS".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1016055 | BD ULTRA FINE PEN NEEDLE | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON DICKINSON AND CO. | 2865 | 0022825 | 10885403928659 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |