FDA Adverse Event Malfunction Summary report: N

VISUAL-ICE CRYOABLATION SYSTEM

MDR report key: 10551168 · Received September 17, 2020

Report

Report Number
2134265-2020-12841
Event Type
Malfunction
Date Received
September 17, 2020
Date of Event
August 18, 2020
Report Date
December 21, 2020
Manufacturer
GALIL MEDICAL INC.
Product Code
GEH
UDI-DI
00867379000006
PMA / PMN Number
K113860
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE VISUAL-ICE SYSTEM WAS RETURNED TO THE DISTRIBUTOR FOR TECHNICAL ANALYSIS AND THE ARGON AND HELIUM SOLENOIDS WERE FOUND FAULTY. THE DISTRIBUTOR FIELD SERVICE ENGINEER REPORTED THAT THEY DO NOT INTEND TO ORDER THE REPLACEMENT PARTS UNTIL JANUARY 2021. THE SYSTEM WILL REMAIN OUT OF SERVICE AT THEIR FACILITY UNTIL SUCH TIME THAT THE FAULTY SOLENOIDS CAN BE REPLACED AND THE SYSTEM RESTORED TO PROPER WORKING ORDER. WITH ALL THE AVAILABLE INFORMATION, BOSTON SCIENTIFIC CONCLUDES THAT THE ROOT CAUSE OF THE REPORTED INSUFFICIENT ICE WAS TRACED TO A COMPONENT FAILURE. H3 OTHER TEXT : DEVICE EVALUATED BY THIRD PARTY DISTRIBUTOR.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A RENAL CRYOABLATION PROCEDURE THE SYSTEM FAILED TO REACH THE DESIRED FREEZING TEMPERATURE. THE ISSUE COULD NOT BE RESOLVED, THEREFORE THE ICE FORMATION WAS SMALLER THAN EXPECTED. THE TREATMENT WAS COMPLETED AFTER FOUR FREEZE CYCLES WITH NO INJURY TO THE PATIENT; HOWEVER, THE INTENDED ABLATION ZONE WAS NOT REACHED. IN THE PHYSICIAN'S OPINION THE TREATMENT WAS UNSUCCESSFUL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A RENAL CRYOABLATION PROCEDURE THE DESIRED FREEZING TEMPERATURE WAS NOT REACHED. THE ISSUE COULD NOT BE RESOLVED, THEREFORE THE ICE FORMATION WAS SMALLER THAN EXPECTED. THE TREATMENT WAS COMPLETED AFTER FOUR FREEZE CYCLES WITH NO INJURY TO THE PATIENT; HOWEVER, THE INTENDED ABLATION ZONE WAS NOT REACHED. IN THE PHYSICIAN'S OPINION THE TREATMENT WAS UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1015192 VISUAL-ICE CRYOABLATION SYSTEM UNIT, CRYOSURGICAL, ACCESSORIES GEH GALIL MEDICAL INC. FPRCH6000 VL0170 00867379000006

Patients

Seq Age Sex Outcome Treatment
1