CANGAROO ENVELOPE
Report
- Report Number
- 3005619880-2020-00045
- Event Type
- Injury
- Date Received
- September 16, 2020
- Date of Event
- July 1, 2020
- Report Date
- September 16, 2020
- Manufacturer
- AZIYO BIOLOGICS INC.
- Product Code
- FTM
- UDI-DI
- 00859389005034
- PMA / PMN Number
- K192616
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
MANUFACTURING REVIEW OF THE CANGAROO ENVELOPE DEVICE HISTORY RECORD FOR THE REPORTED LOT REFLECTS THAT ALL PACKAGED AND LABELED UNITS WERE QUALITY RELEASED TO DISTRIBUTION ON 2/12/2020 HAVING MET ALL INTERNAL QC ACCEPTANCE REQUIREMENTS. ALL STERILIZATION PROCESSING RECORDS AND BIOBURDEN TESTING RESULTS INDICATE A SUCCESSFUL STERILIZATION CYCLE, AND PASSING RESULTS OF LAL AND STERILITY BIS ALLOWED THE SUBASSEMBLY LOT TO BE RELEASED FOR FURTHER PACKAGING AND LABELING ACTIVITIES. THERE WERE NO NON-CONFORMANCES DURING MANUFACTURING OR STERILIZATION POTENTIALLY IMPACTING THE FINAL ACCEPTANCE OF THIS MANUFACTURING LOT. IN LIEU OF A REQUEST FOR THE OEM SUPPLIER DHR REVIEW, IT IS NOTED THAT AZIYO PROCESSES THE NON-STERILE ENVELOPE MATERIALS INCLUDING CUTTING, SUTURING, PACKAGING AND STERILIZING THE PRODUCT. IT IS ALSO NOTED THAT PER THE INSTRUCTIONS FOR USE (ART-20662) PROVIDED WITH THE PACKAGED CANGAROO ENVELOPE DEVICE, THAT INFECTION IS LISTED AS A POTENTIAL COMPLICATION ASSOCIATED WITH THE SURGICAL PROCEDURE AND USAGE OF THE CANGAROO ENVELOPE. ALTHOUGH THE EXACT CAUSE OF THE REPORTED INFECTION CANNOT BE CONCLUSIVELY DETERMINED, INFECTION IS A KNOWN COMPLICATION ASSOCIATED WITH THE SURGICAL IMPLANTATION OF THE CIED AND THE CANGAROO ENVELOPE DEVICES. SHOULD AZIYO RECEIVE ANY ADDITIONAL DETAILS RELATED TO THIS EVENT, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED BY AZIYO BIOLOGICS' BUSINESS PARTNER BOSTON SCIENTIFIC, THAT A CANGAROO ENVELOPE (MODEL CMCV-009-MED LOT# M20B1068) WAS USED WITH A BOSTON SCIENTIFIC PACEMAKER (MODEL #/SERIAL # UNKNOWN) AND IMPLANTED ON (B)(6) 2020. IT WAS REPORTED THAT THE PATIENT WAS ADMITTED ON (B)(6) 2020 DUE TO INFECTED PACEMAKER WITH WOUND DEHISCENCE. THE ENTIRE SYSTEM, INCLUDING THE AZIYO CANGAROO ENVELOPE WERE EXPLANTED ON 7/1/2020. WOUND WAS IRRIGATED WITH ANTIBIOTIC SOLUTION AFTER EXPLANTING THE DEVICE. THE ISSUE WAS REPORTED TO AZIYO BIOLOGICS ON 8/18/2020.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1003910 | CANGAROO ENVELOPE | MESH, SURGICAL - FTM | FTM | AZIYO BIOLOGICS INC. | CMCV-009-MED | M20B1068 | 00859389005034 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | BSC PACEMAKER (MODEL#/SERIAL# - UNKNOWN) |