FDA Adverse Event Injury Summary report: N

CANGAROO ENVELOPE

MDR report key: 10541396 · Received September 16, 2020

Report

Report Number
3005619880-2020-00045
Event Type
Injury
Date Received
September 16, 2020
Date of Event
July 1, 2020
Report Date
September 16, 2020
Manufacturer
AZIYO BIOLOGICS INC.
Product Code
FTM
UDI-DI
00859389005034
PMA / PMN Number
K192616
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURING REVIEW OF THE CANGAROO ENVELOPE DEVICE HISTORY RECORD FOR THE REPORTED LOT REFLECTS THAT ALL PACKAGED AND LABELED UNITS WERE QUALITY RELEASED TO DISTRIBUTION ON 2/12/2020 HAVING MET ALL INTERNAL QC ACCEPTANCE REQUIREMENTS. ALL STERILIZATION PROCESSING RECORDS AND BIOBURDEN TESTING RESULTS INDICATE A SUCCESSFUL STERILIZATION CYCLE, AND PASSING RESULTS OF LAL AND STERILITY BIS ALLOWED THE SUBASSEMBLY LOT TO BE RELEASED FOR FURTHER PACKAGING AND LABELING ACTIVITIES. THERE WERE NO NON-CONFORMANCES DURING MANUFACTURING OR STERILIZATION POTENTIALLY IMPACTING THE FINAL ACCEPTANCE OF THIS MANUFACTURING LOT. IN LIEU OF A REQUEST FOR THE OEM SUPPLIER DHR REVIEW, IT IS NOTED THAT AZIYO PROCESSES THE NON-STERILE ENVELOPE MATERIALS INCLUDING CUTTING, SUTURING, PACKAGING AND STERILIZING THE PRODUCT. IT IS ALSO NOTED THAT PER THE INSTRUCTIONS FOR USE (ART-20662) PROVIDED WITH THE PACKAGED CANGAROO ENVELOPE DEVICE, THAT INFECTION IS LISTED AS A POTENTIAL COMPLICATION ASSOCIATED WITH THE SURGICAL PROCEDURE AND USAGE OF THE CANGAROO ENVELOPE. ALTHOUGH THE EXACT CAUSE OF THE REPORTED INFECTION CANNOT BE CONCLUSIVELY DETERMINED, INFECTION IS A KNOWN COMPLICATION ASSOCIATED WITH THE SURGICAL IMPLANTATION OF THE CIED AND THE CANGAROO ENVELOPE DEVICES. SHOULD AZIYO RECEIVE ANY ADDITIONAL DETAILS RELATED TO THIS EVENT, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED BY AZIYO BIOLOGICS' BUSINESS PARTNER BOSTON SCIENTIFIC, THAT A CANGAROO ENVELOPE (MODEL CMCV-009-MED LOT# M20B1068) WAS USED WITH A BOSTON SCIENTIFIC PACEMAKER (MODEL #/SERIAL # UNKNOWN) AND IMPLANTED ON (B)(6) 2020. IT WAS REPORTED THAT THE PATIENT WAS ADMITTED ON (B)(6) 2020 DUE TO INFECTED PACEMAKER WITH WOUND DEHISCENCE. THE ENTIRE SYSTEM, INCLUDING THE AZIYO CANGAROO ENVELOPE WERE EXPLANTED ON 7/1/2020. WOUND WAS IRRIGATED WITH ANTIBIOTIC SOLUTION AFTER EXPLANTING THE DEVICE. THE ISSUE WAS REPORTED TO AZIYO BIOLOGICS ON 8/18/2020.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1003910 CANGAROO ENVELOPE MESH, SURGICAL - FTM FTM AZIYO BIOLOGICS INC. CMCV-009-MED M20B1068 00859389005034

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention BSC PACEMAKER (MODEL#/SERIAL# - UNKNOWN)