DR FOG TREATED SPONGE, STERILE
Report
- Report Number
- 1836161-2020-00043
- Event Type
- Malfunction
- Date Received
- September 16, 2020
- Date of Event
- August 21, 2020
- Report Date
- August 21, 2020
- Manufacturer
- ASPEN SURGICAL PRODUCTS, CALEDONIA
- Product Code
- OCT
- PMA / PMN Number
- K932449
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
ASPEN SURGICAL RECEIVED A REPORT FROM THE DISTRIBUTOR INDICATING THAT PRODUCT WAS FOUND WITH SEAL ISSUES. THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. THE MANUFACTURING LOT NUMBER ALONG WITH SAMPLES WERE PROVIDED FOR REVIEW. THE DISTRIBUTOR INDICATED THAT THE DEFECTS WERE FOUND DURING INCOMING INSPECTION. A REVIEW OF THE SAMPLES CONFIRMED THE ISSUE FROM THE DISTRIBUTOR. ANALYSIS OF THE FINISHED GOOD LOT NUMBER WAS REVIEWED. NO NON-CONFORMANCE'S WERE NOTED DURING THE MANUFACTURING PROCESS. ACCORDING TO THE MANUFACTURING PROCESS, THESE PARTS ARE MANUALLY LOADED INTO RECESSED POCKETS ON A PACKAGING MACHINE PER SPECIFIED WORK INSTRUCTIONS. IF INCORRECTLY LOADED OR NOT ORIENTATED PROPERLY, THE PRODUCT INTERFERES WITH THE SEALING PROCESS. THEREFORE A LIKELY ROOT CAUSE FOR THE DEFECT MAY BE ATTRIBUTED TO AN OPERATOR ERROR. THE IFU WHICH IS RECEIVED WITH THE PRODUCT, ALONG WITH THE POUCH LABEL, IDENTIFIES THIS FAILURE MODE WITH THE SYMBOL "DO NOT USE IF PACKAGE IS DAMAGED". THIS INDICATES THAT THE DEVICE SHOULD NOT BE USED IF THE PRODUCTS STERILE BARRIER SYSTEM OR ITS PACKAGING IS COMPROMISED. ADDITIONALLY, PRODUCTION SUPERVISORS AND OPERATIONS WERE NOTIFIED OF THIS ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
ASPEN SURGICAL RECEIVED A REPORT FROM THE DISTRIBUTOR INDICATING THAT A DR FOG SPONGE WAS DISCOVERED WITH A SEALING ISSUE. THE ITEM WAS NOT IN USE. NO INJURY/DEATH WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM UNDER MULTIPLE LOTS FOR ISSUE. LOT NUMBERS AND ASSOCIATED RECORDS ARE AS FOLLOWS: C-1181085, LOT 215132, C-1181086, LOT 215905, C-1181087, LOT 218690, C-1181090, LOT 219227.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1005439 | DR FOG TREATED SPONGE, STERILE | ANTI-FOG | OCT | ASPEN SURGICAL PRODUCTS, CALEDONIA | DF-3120 | SEE ABOVE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |